Companion Portable Technical Service Manual • PN 20562220 Rev B
8
Companion P
or
tables
III
Safety
Table 4*
Guidance and manufacturers declaration—electromagnetic immunity
The Companion 1000/T is intended for use in the electromagnetic environment specified below. The customer or the user of the
Companion 1000/T should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level Electromagnetic environment—guidance
Conducted RF
3Vrms
Not Applicable
Portable and mobile RF communications equipment should
IEC 61000-4-6
150kHz to 80 MHz
Battery powered
be used no closer to any part of the Companion 1000/T,
device
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency
of the transmitter.
Recommended separation distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2,5 GHz
where P is the maximum output power rating of the
Radiated RF
3 V/m
3 V/m
transmitter in watts (W) according to the transmitter
IEC 61000-4-3
80 MHz to 2,5 GHz
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survery
a
, should be less than
the compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, ama-
teur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To asses the electromag-
netic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the Companion 1000/T is used exceeds the applicable RF compliance level above, the Companion 1000/T
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the Companion 1000/T.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
* This table is included as a standard requirement for equipment which has been tested to specific test levels and over specific