Safety Features
This mixer includes the following safety features. Please reference Features and Operation section and Warnings and
Notifications section for further details.
Quick Stop
Press to immediately stop operation in the event of an unforeseen safety risk to the operator or environment.
Remote Control Enabled Notification
When remote control is enabled, notification appears on screen when attempting local operation.
Resume on Power Up Notification
When mixer is programmed to Resume on Power Up, a visual and audible warning occurs during a 5 second start up delay.
See Feature Settings for additional information.
Overheat Warning
Operation stops when control board temperature reaches 70 degrees C. Notification and instructions appear on screen.
Overheat Motor Warning
Operation stops when motor temperature reaches 105 degrees C. Notification and instructions appear on screen.
Overload Warning
Operation stops when torque requirements exceed capabilities or output shaft is locked. Notification and instructions appears
on screen.
Chuck Shaft Guard
Provides guarding of rotating chuck and mixing shaft.
Thermal Cut Off Switch
If the mixer temperature (at base of motor) exceeds 60 degrees C, an internal thermal cutoff is activated eliminating all
power to the mixer and display will be blank. Once mixer cools to below 40 degrees C, power is automatically reinstated
with mixer in the paused state. Use keypad to continue operation.
Note
: If Resume on Power Up feature is active, operation will resume immediately when power is restored.