2
Safety Instructions
CAUTION / WARNING
1.
All operators must be familiar with the mixer and read this entire manual. If the mixer is used in a manner not specified
by the manufacturer, the protection provided by the equipment may be impaired.
2. Ensure that only trained staff operate the mixer.
3. Because the combinations of products, impellers, stirring vessel, and medium are nearly endless, user safety cannot
be ensured solely with product design requirements. For this reason, it may become necessary for users to take
additional precautionary safety measures.
4 Ensure the Power Switch and Quick Stop switch of the mixer can be accessed immediately, directly and without risk at
any time, and above all, in case of fault.
5. The mixer must be securely fixed to a stable support. The high torque produced by the mixer requires particular care
in the choice of support for the mixer and vessel. Mixer not intended for hand held operation.
6. All mixing impellers must be in good condition with straight shafts. If the mixer vibrates at high speeds check the shaft
for damage and repair or replace it.
7. The mixer must be supplied with the rated voltage (see serial label).
8. Ensure that the mixing impeller does not contact the stirring vessel.
9. Spinning impellers can cause severe personal injuries. Operators must use extreme care and good judgment when
mixing at any speed.
10. Always ensure the power switch in the OFF position or disconnect the power before changing the impeller, attaching
the shaft guard, and fitting accessories. Do not attempt to service or reposition the mixer while shaft/impeller is
rotating.
11.
CAUTION
: This is not an explosion proof mixer. Do not use with highly flammable or explosive materials. Do not
operate the mixer in explosive atmospheres, with hazardous substances or under water.
12. Extreme care must be taken when mixing chemicals so that no chemicals are splashed outside the mixing vessel.
Care must be taken when changing to faster mixing speeds. On power up, the mixer will display its set speed and will
climb to that speed when the PAUSE/PLAY button is pushed. Always start at lowest speed if unsure of maximum safe
speed.
13. Do not operate while standing in water. Keep the mixer dry and do not immerse any part, except the impeller into any
liquids. Protect from splashing.
14. Ensure that no loose clothing, jewelry, or hair can become entangled in any rotating parts. A fast spinning chuck can
cause injury to an operator.
15 Power can be interrupted to the mixer by pressing the power switch or by disconnecting the mains cord. If rotating, this
will cause the mixer to stop.
16. Shaft rotation can also be stopped by pressing the PAUSE/PLAY button. This does not disconnect power to the mixers
internal drive circuit.
17. Wear safety goggles and suitable clothing when operating the mixer.
18. Repairs must be carried out only by technicians authorized by Caframo.
19. The motor, transmission or housing surface may become hot during operation.
20. Motor fan cover must be secured in place during operation.
21. Always confirm the direction of rotation as noted on the Home Screen dashboard prior to operation.
22. Ensure motor has been stopped by pressing the PAUSE/PLAY button before turning power switch off. Do not use the
power switch to stop rotation.
23. A Quick Stop switch has been incorporated for safety. Pressing the Quick Stop will immediately stop operation.
24. Mixer must be used with supplied power cord. The cord and grounded plug type are important safety elements. Do not
use an inadequately rated cord or ungrounded plug. Always connect the mixer to a grounded outlet. Do not alter the
plug.
25. Following periods of operation, ensure mixer has adequately cooled before handling.
26. Noise levels may exceed 80 decibels under certain operating conditions. Ear protection recommended.
27. Only use guarding provided by the manufacturer (see accessories section).
28. If the reason for loss of power is uncertain, switch off the mixer and check for mechanical hazards prior to resuming
operation. When Resume on Power Up feature is set to on, a 5 second delay occurs, 5 short beeps sound and a
notification appears on screen prior to start. When audible alarm sounds, please keep clear of moving components. In
the absence of audible alarm or on screen notification, switch off the mixer and check for mechanical hazards.