CAEN
Electronic
Instrumentation
DT54xx
USB
High
Voltage
Power
Supply
13
4.
USB
Communication
Protocol
The
following
Protocol
allows
to
communicate
with
the
module.
The
Protocol
is
based
on
commands
made
of
ASCII
characters
strings.
The
recognized
values
are
as
follows:
Version
DT5472
DT5471
DT5470
Parameter
unit
VSRES
V
0.1
0.5
1
VMRES
V
0.1
0.5
1
RAMPMAX
V/s
500
500
500
RAMPMIN
V/s
1
1
1
TRIPMAX
s
1000
1000
1000
IMAX
µA
1050
510
210
ISRES
µA
0.2
0.1
0.04
IMRESH
µA
0.2
0.1
0.04
IMAXL
µA
105
51
21
IMRESL
µA
0.02
0.01
0.004
Command
Format
string
$CMD:***,PAR:***,VAL:xxx.xx<CR,LF>
CMD
:
SET,
MON
PAR
:
VSET,
ISET,
ON
..
VAL
:
value
Format
of
response
string
#CMD:***,VAL:xxx.xx<CR,LF>
CMD
:
OK,
ERR.
VAL
:
value
All
fields
are
in
ASCII
format
Resolution
values
are
passed
as
real
numbers
Monitor
Commands
$CMD:MON,PAR:BDNAME<CR,LF>
Read
out
module
name
DT5472
500V/1mA
DT5471
3kV/500uA
DT5470
5kV/200uA
$CMD:MON,PAR:TYPE<CR,LF>
Read
out
the
channel
type:
0:
500V/1mA
1:
3kV/500uA
2:
5kV/200uA
$CMD:MON,PAR:POLARITY<CR,LF>
Read
out
channel
polarity.
$CMD:MON,PAR:VSET<CR,LF>
Read
out
VSET
value