PreDAC User’s Manual
28
2.3.5
TEMP command
TEMP Command (“TEMP:?
\r\n
”
) allows user to read temperature from the
internal temperature sensor. Temperature value is updated every 10 seconds.
If temperature rises over the threshold of 50°C, the over-temperature fault trips
and the output voltages are set to 0 V level. After a fault event occurs it is necessary to
reset the internal status register in order to be able to reactivate voltage outputs.
The reply to the TEMP command is in the format “TEMP:
value
\r\n
”, where
value
is the integer temperature readback value (in [°C]).
Example:
TEMP
read example:
TEMP:?
\r\n
TEMP:28
\r\n
Summary of Contents for PreDAC