25
Setup
Optional: Insert the CO
2
Canister (Prehospital Only)
Some SCEs include the simulation of CO
2
exhalation. The following instructions show how to
safely connect the CO
2
canister to the simulator.
WARNING
: Careful handling, including the use of eye protection, is required when using CO
2
canisters.
Please read and understand all the important cautions and warnings on removing canisters as
well as safety steps that must be used when handling CO
2
canisters.
Use of CO
2
Canisters
Store the CO
•
2
canisters in a dry location between 32° and 104° F. (0° to 40°C)
Do not expose the CO
•
2
canister to heat above 140° F, as rupture may occur.
Never point the CO
•
2
canister toward your face or someone nearby.
Use only CAE Healthcare specified CO
•
2
canisters.
Do not remove the canister from the regulator base until empty. The canister end is
•
punctured when screwed into the regulator base.
Never ship the CO
•
2
canister attached to the regulator assembly.
Assembly of the CO
2
Regulator
Care must always be taken when using high-pressure equipment.
•
Do not disassemble or alter the regulator.
•
Dry completely if the regulator becomes wet.
•
Discontinue use of this equipment if leakage or visible damage is evident.
•