BYK-Gardner GmbH
Appendix | 8
300 100 114 E 2103
micro-TRI-color
51
8.3.4 Cleaning the Standard Tiles
Using dirty or damaged standard tiles can impact the accuracy of measurements
significantly. Clean the standard tiles when they are dirty in order to avoid measur-
ing inaccuracy.
NOTICE
Be careful when cleaning the standard tiles
Apply only slight pressure as you clean and make certain there are no large parti-
cles stuck in the cloth that could damage the surface. We highly recommend to
handle the standard tiles with great care. Store them always enclosed.
Use a new lint-free cloth, dust-free lens paper or an optical cloth for cleaning. For
dirt that is difficult to remove, use an optical cloth dipped in propanol. Then wipe
the surface with a dry optical cloth.
8.3.5 Calibration and Repair Service
Our Service Offer
BYK-Gardner´s global network of own ISO/IEC 17025 accredited service points is
equipped with the full line of reference measuring standards and tools needed to
ensure highest quality service on a global basis:
• Preventive maintenance for a longer life
• Certification services for standards and physical testing tools
• Repair service
• On-site service
You can find more details on our website:
•
https://www.byk-instruments.com/tw/en/technical-services
Preventive Maintenance
To increase the reliability and life time of your instrument, regular inspections and
optimizations are recommended. With our preventive maintenance solution your
instrument will always be in the best shape. We clean the optics, check all func-
tions, test and, if required, adjust the measured values by using reference stan-
dards.
Preventive maintenance is recommended every 12 – 24 months depending on us-
age and should only be performed by BYK-Gardner technical service centers. For
the list of certified service centers see
.
Summary of Contents for micro-TRI-color
Page 1: ...Measure what you see micro TRI color Type 345 000 015 Operating Instructions...
Page 32: ...6 Browse BYK Gardner GmbH 32 micro TRI color 300 100 114 E 2103...
Page 46: ...8 Appendix BYK Gardner GmbH 46 micro TRI color 300 100 114 E 2103...
Page 53: ...BYK Gardner GmbH 300 100 114 E 2103 micro TRI color 53 Notes...
Page 54: ...BYK Gardner GmbH 54 micro TRI color 300 100 114 E 2103...
Page 55: ...BYK Gardner GmbH 300 100 114 E 2103 micro TRI color 55...