BYK-Gardner GmbH
Calibration and Check | 4
300 100 114 E 2103
micro-TRI-color
23
4 Calibration and Check
In order to work properly, the instrument needs to be calibrated on a white stan-
dard and should be frequently monitored on a green checking standard:
•
White calibration:
Necessary every
30
days.
•
Green check:
Recommended every
7
days.
As long as the white calibration is valid, the instrument displays a green check-
mark. The green checkmark is placed on the calibration icon in the main menu.
The instrument automatically indicates when white calibration is necessary. The in-
dication is a red cross instead of a green checkmark.
See also
2
4.1 Perform White Calibration
A white calibration is necessary. This is indicated by a red cross, placed on the cali-
bration icon in the main menu.
1. Open the snap closure.
2. Click on the calibration icon
in the main menu and place the in-
strument on the
white
tile as you see below.
Illustration 17:
Instrument placed on white calibration tile
3. The screen below is displayed.