3.6.3 Menu: SETTINGS / CONDUCTIVITY
Settings for permeate conductivity:
D
The main selection menu appears with
“Start“
being activated.
St
a
r
t
START
S
e
t
t
i
n
g
s
START
Fig. 15:
Main selection menu: “Start“
,
“Settings“
D
The menu item
“
z
CONDUCTIVITY“
is active.
1. Menu level (selection menu “SETTINGS“)
z
GENERAL
General
device settings
z
PRESSURE
Pressure of the permeate
z
CONDUCTIVITY
Conductivity value
(ACTIVE)
z
WCF SETUP
Permeate output, WCF menu
Fig. 16:
Menu Level 1: “Settings“
Press the
[
]
key
again to set the limits for the alarm and
the
activation/deactivation
of the
conductivity alarm
.
2. Menu level (adjustable parameters)
X
CONDUCTIVITY LIMIT
Conductivity limit value
(ACTIVE)
X
CONDUCTIVITY ALARM
Conductivity alarm
Fig. 17:
Menu Level 2: “Settings/CONDUCTIVITY“
Note:
In case you activate the
“STOP“
setting,
you have to deactivate and then activate the
device again.
Note:
It is not possible to set the last digit “10
0
“
at first. Please observe the right input order.
Perform the settings of the
“CONDUCTIVITY“
selection menu in accordance with
Fig. 18
.
START
z
GENERAL
z
PRESSURE
z
CONDUCTIVITY
z
WCF SETUP
X
CONDUCTIVITY LIMIT
X
CONDUCTIVITY ALARM
0
90 μS
Conductivity limit
0
90 μS
Conductivity limit
0
5
0 μS
Conductivity limit
0
5
0 μS
Conductivity limit
0
5
0 μS
Conductivity limit
START
X
CONDUCTIVITY LIMIT
X
CONDUCTIVITY ALARM
*
DISABLE
ALARM
Conductivity alarm
*
DISABLE
*
ALARM
Conductivity alarm
DISABLE
*
ALARM
Conductivity alarm
Fig. 18:
Programming procedure: “CONDUCTIVITY“
1x