Table 18. Color + Ophthalmic
Index Label
MI
TIS
TIB
TIC
Scan
Non-Scan
Non-
Scan
A
aprt
≤1
cm
2
A
aprt
>1
cm
2
Maximum Index Value
0.162
0.00145
0.0167
-
0.0477
0.0340
Assoc Acoustic P
ar
ameter
Pr.3
(MPa)
0.355
W
o
(mW)
0.0633
0.719
0.719
0.719
min of [W
.3
(z
1
),
I
TA.3
(z
1
)]
(mW)
-
z
1
(cm)
-
z
bp
(cm)
-
z
sp
(cm)
1.65
1.22
d
eq
(Z
sp
)
(cm)
0.0358
f
c
(MHz)
4.80
4.80
4.89
-
4.89
4.89
Dim of A
aprt
X (cm)
0.65
0.5
-
-
0.5
Y (cm)
0.65
0.5
-
-
0.5
Other Inf
or
mation
PD
(μsec)
0.175
PRF
(Hz)
1100
p
r
@PII
max
(MPa)
0.507
d
eq
@PII
max
(cm)
0.311
Focal Length
FLx (cm)
4.0
5.0
-
FLy (cm)
4.0
lnf
-
I
PA.3
@MI
max
(W/cm2)
6.48
Operating Con-
trol Conditions
Ophthalmic
✓
✓
✓
Note 1:
Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
Note 2:
Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or neonatal
cephalic uses.
Note 3:
Information on MI and TI need not be provided if the equipment meets both the exemption clauses given in 51.2aa) and 51.2
dd).
Acoustic Output
Specifications
53