Cleaning and Disinfecting the Probe
WARNING!
Failure to disinfect the probe may result in an increased spread of pathogens.
CAUTION!
Only clean the probe with approved cleaning products and wipes. Improper cleaning or disinfec-
tion methods or using non-approved cleaning and disinfecting solutions may damage equipment.
This section provides information and instructions for properly cleaning and disinfecting the Butterfly iQ™ probe. Fol-
lowing these instructions will also help to avoid damaging the probe during cleaning and disinfection. After each ex-
am, clean and disinfect the Butterfly iQ™.
While the Cleaning and Disinfection guidance contained here has been validated for effectiveness, a list of cleaning
and disinfection products that are compatible with the Butterfly iQ™ probe, but not tested for effectiveness by Butter-
fly, can be found in the "Compatible Cleaning and Disinfection Products" article available at
. The products listed in the Compatible Cleaning and Disinfection Products article will
not impact the functionality of the probe when used according to the instructions provided by the manufacturer of the
product.
Cleaning the Probe
CAUTIONS!
• Prevent any fluid from entering electrical or metal portions of the cable’s connector during the
cleaning and disinfecting process. Damage due to fluid in these areas may result.
• Prevent any fluid from splashing on your mobile device’s touchscreen during scanning and dur-
ing cleaning. Damage due to fluid may result.
To clean the probe:
1. After each use of the probe, use one of the recommended liquid saturated wipes (Super Sani-Cloth® Germicidal
Disposable Wipes by PDI, Inc., Super Sani-Cloth® AF3 Disposable Wipes by PDI, Inc., or a lint-free cloth mois-
tened with water) to remove ultrasound transmission gel from the probe.
2. Disconnect the probe from the mobile device.
3. Wipe the probe, strain relief, cable, and connector with one of the recommended liquid saturated wipes for one
(1) minute and until visibly clean.
4. Change the wipes as necessary and repeat the above step until the probe is visibly clean.
5. To dry the probe, use a soft cloth, blot the lens dry. Do not wipe the lens. Dry the rest of the probe, cable, strain
relief, and connector.
6. Visually inspect the probe in a well-lit area to ensure all surfaces are clean. If the probe is not clean, repeat the
cleaning steps above.
7. Dispose of cleaning material in accordance with all applicable regulations.
For the most up to date list of approved cleaners please visit
Cleaning and Disinfecting the Probe
34
Maintenance