RESISTOMAT
®
Model 2302-V001
Page 40
Introduction &
applications
Description of functions
Brief guide
Operation
Calibration
Test and offset
instructions
RS232 data
output
Accessories and Circuit diagram
RS232 data
output
Parameter 9:
This parameter contains the table header for printouts. The header must
not exceed 70 characters or contain special German characters.
Parameter 10:
SRQ flag (service request).
Depending on its value, this parameter effects a continuous printer
output on every new A/D conversion, or only when a service
request signal ( input/output line P5 ) is present.
Value:
Meaning:
0
Continuous printout
>1
Printout on SRQ signal
( Default setting: 0 )
Parameters 11 ... 17: These parameters cannnot be accessed.
Parameter 18:
Baud rate
This parameter contains the software-configured baud rate
Value:
Meaning:
1
1200
baud
2
2400
baud
3
4800
baud
4
9600
baud
5
19200
baud
( Default setting: 4 )
Parameter 19:
Printer / interface operation
During software configuration, this parameter selects printer or
interface operation.
Value:
Meaning:
0
Interface operation
1
Printer operation
( Default setting: 0 )
Parameter 20:
Single/double addressing
During software configuration, this parameter selects single or double
addressing.
Value:
Meaning:
0
Single addressing
1
Double addressing
( Default setting: 0)
Parameters 21 ... 49: These parameters are not used and cannot be written or read.
Parameter 50:
Header
This parameter contains a header for every printed page. The
header must not exceed 55 characters or contain special German
characters.