Trouble-shooting table aid
Problem
Possible causes
Solution
Hand switch / drive
system not functioning
n
Main power cable is not
plugged in
n
No power supply to socket
n
Hand switch plug not properly
connected
n
Hand switch or drive system
defective
n
Functions are blocked on
hand switch
F
Insert main power cable
F
Check socket/fuse box
F
Test the plug connection
F
Inform your operator about
necessary repairs
F
Unlock functions (see
page 24)
drives only operates for
a short time when
buttons are pressed
n
Too much weight on the bed
n
Bed is blocked by an obstacle
F
Reduce load
F
Remove obstacle
Individual drives operate
in one direction only
n
defective hand switch, motor
drives or control unit
F
Inform your operator about
necessary repairs
The wrong function is
active when hand switch
buttons are pressed
n
Internal motor plugs
incorrectly connected (mixed up)
F
Inform your operator about
necessary repairs
drives stop suddenly
after prolonged adjust-
ment times
n
Thermo switch in transformer
(drive unit) is triggered on
F
Pay attention minimum pause
(see page 22)
Mattress base not
horizontal
n
Lifting actuators not running
synchronously
F
Raise or lower mattress base
to highest or lowest setting at
least once a day.