67
Adjustmentandstart-up
9.14 Knowing the Test menu
9.14.1 Modifying the Test menu access code
See chpt. 9.9 to access the Test menu.
System
Code
0***
Confirm code
0***
Test
Enter the new
code
Confirm the new
code
If the default code (0000) is entered, the code will not be requested to access the menu.
9.14.2 checking the outputs functions
See chpt. 9.9 to access the Test menu.
• Make sure that the mode "Hold" is deactivated. Refer to chpt. 9.12.1.
• The
T
icon is displayed in place of the
icon as soon as the check for the correct working of an
output has started. During the check the related output does not react according to the measured
physical value.
Test
Outputs
AC2:
AC1:
INPUT
INPUT
TR1:
OFF/ON
TR2:
OFF/ON
AC1
: check that current output 1 is working correctly by entering a current value and then selecting "OK".
AC2
: check that current output 2 is working correctly by entering a current value and then selecting "OK".
TR1
: check that transistor output 1 is working correctly by selecting the status of the transistor ("ON" or "OFF") then "OK".
TR2
: check that transistor output 2 is working correctly by selecting the status of the transistor ("ON" or "OFF") then "OK".
9.14.3 checking the outputs behaviour
See chpt. 9.9 to access the Test menu.
• Make sure that the mode "Hold" is deactivated. Refer to chpt. 9.12.1.
• The
T
icon is displayed in place of the
icon as soon as the check for the correct working of an
output has started. During the check the related output does not react according to the measured
physical value.
The feature allows for simulating the measurement of the process value to check if the outputs are correctly configured.
Test
Value:
Sensor
INPUT
PVar:
pH
TempC
mV_pH
mV_ORP
TempF
PVAR
: choose the process value to be tested.
VALUE
: enter a process value selected from the "PVAR" function above to check output behaviour.
English