Adjustmentandstart-up
9.11.5 setting the data displayed in the process level
See chpt. 9.9 to access the Parameters menu.
Warning
risk of injury due to wrong adjustment.
▶
Before setting the parameters for the display, choose the type of probe mounted on the device. Refer to
chpt. 9.11.11.
Display
Param
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
pH
TempC
Line1 / Line2:
Filter:
None
Fast
Slow
Enabled
Line1 / Line2:
PVar:
Disabled
mV_pH
mV_ORP
TempF
pH
°C
Unit:
mV
°F
If PVar = "TempC"
If PVar = "TempF"
If PVar = "mV_pH" or "mV_ORP"
If PVar = "pH"
PVAR
: choose the process value to be displayed in the line selected.
UNIT
: choose the unit for the process value displayed.
FILTER
: choose the filter level for the measurement values displayed on the line selected. Three filter levels are
proposed: "slow", "fast" or "none".
"slow"
"fast"
"none"
150 ms
6 s
t
30 s
t
t
Fig. 36: Filter curves
English
Type 8202 ELEMENT