35
Adjustmentandstart-up
9
ADjusTMENT AND sTArT-up
• The settings can only be done on a device with a display module.
• Do not remove the display module while making the settings on the device.
9.1
safety instructions
Warning
risk of injury due to non-conforming adjustment.
Non-conforming adjustment could lead to injuries and damage the device and its surroundings.
▶
The operators in charge of adjustment must have read and understood the contents of the Operating
Instructions.
▶
In particular, observe the safety recommendations and intended use.
▶
The device/installation must only be adjusted by suitably trained staff.
Warning
Danger due to non-conforming start-up.
Non-conforming start-up could lead to injuries and damage the device and its surroundings.
▶
Before start-up, calibrate the pH/Redox ("ORP") sensor. Refer to chpt. 9.12.4.
▶
Before start-up, make sure that the staff in charge have read and fully understood the contents of the Operat-
ing Instructions.
▶
In particular, observe the safety recommendations and intended use.
▶
The device/installation must only be commissioned by suitably trained staff.
9.2
Knowing the operating levels
The device has 2 operating levels:
process level
This level is used:
• to read the measured values of 2 measurable variables selected in the Parameters menu,
• to read both the lowest and highest values of the chosen measurable variable, that have been measured by the
device since the power-up of the device or since the latest reset (this feature is not active by default),
• to reset both the lowest and highest values of the chosen process value, if the feature has been activated,
• to read the current values emitted on the 4...20 mA outputs.
• to get the state of the device and of the pH/Redox ("ORP") sensor with the icons.
English
Type 8202 ELEMENT