Type MS08
General Settings - Parameter
61
Name
Data type
Unit SI
Range
Event timer
(ms)
Inhibit time
(ms)
TOCeq
(Total Organic Car-
bon)
REAL32
g/l
0…880
5000
500
Tab. 37: PDO 3 - Transmitted data and their default values
Multiplexed PDO
Name
Data type
Unit SI
Range
Event timer
(ms)
Inhibit time
(ms)
Transmission 254nm
REAL32
%
3…98.8
-
500
Transmission 530nm
REAL32
%
3…98.8
-
500
Absorption 254nm
REAL32
1/m
0.1…1500
-
500
Absorption 530nm
REAL32
1/m
0.1…500
-
500
SQI (Quality Index)
REAL32
-
500
Tab. 38: Multiplexed PDO - Transmitted data and their default values
Received PDO
The product receives from other participants to the fieldbus the PDO 1 described in the following table:
Name
Data type
Unit SI
Range
Event timer
(ms)
Inhibit time
(ms)
Hold
UNSIGNED8
-
-
-
-
Tab. 39: Received PDO 1 and its default values
To set the transmission time between 2 values of the PDO, do the following procedure:
Make sure that the login user level is
Installer
. Refer to chapter
.
Go to device
SAC254 Sensor
> function
General settings
> tab
Parameter
.
Select
PDO configuration
>
PDO 1
,
PDO 2
,
PDO 3
or
Multiplexed PDO
.
Select
Inhibit time
and enter a value in ms.
Repeat for any value you want to adjust.
To reset all values to default, select
Reset to default values
.
To apply and save the changes, select
Apply and Save
.
The PDO inhibit times are changed.
12.19
Restore all PDOs to their default values
Make sure that the login user level is
Installer
. Refer to chapter
.
Go to device
SAC254 Sensor
> function
General settings
> tab
Parameter
.
Select
PDO configuration
.
Select
Reset to default values
.
The PDOs are set to their default values.