Type MS08
Troubleshooting
79
16.1.4.4
Message
Deviation from default settings detected, resets to default settings
Product status symbol
Possible cause
The default settings have been modified. They will be overwritten by
the ME63 interface.
What to do?
No action.
16.1.4.5
Message
Last calibration is older than %1 days. Recalibrate your device
Product status symbol
Possible cause
The calibration date is due.
What to do?
Recalibrate the sensor. Refer to chapter
16.1.4.6
Message
Sensor error (invalid baseline). Contact the service for your device
Product status symbol
Possible cause
After performing a calibration. The base line is wrong.
What to do?
Clean the sensor.
Check the calibration solution (DI water).
Recalibrate the sensor. Refer to chapter
16.1.4.7
Message
Sensor error (optical path length too short). Contact the service for your
device
Product status symbol
Possible cause
The path length is not the good one.
What to do?
Clean the sensor.
Check the sensor range according to the process.
16.1.4.8
Message
Sensor error (optical path length too long). Contact the service for your
device
Product status symbol
Possible cause
The path length is not the good one.
What to do?
Clean the sensor.
Check the sensor range according to the process.