Operation
At least 250mV of sample
voltage is required in order to
calibrate channels. Note that
channels configured for
millivolt, degree or virtual
input types cannot be
calibrated.
CALIBRATION
From the main menu, press
CONFIG
then
CALIBRATE
.
Find the desired channel using
the scroll buttons as needed then
press the channel to select it.
Press CALIB to show the
calibration screen for the
selected channel.
Use the keypad to enter the
actual value as read from the
local meter, then press
ENTER
.
For diagnostic purposes, the raw sample voltage (or
virtual channel value) will show below the meter window.
Calibration can only be performed on the local site.
SYSTEM MENU
Press
CONFIG
from the main menu to select the
CONFIGURATION
sub-menu then press
SYSTEM
.
CLOCK
Press
CLOCK
to display the time and date settings.
Press
DATE
,
TIME
, or
TIME ZONE
to change the value
shown.
Press the field to be changed then use the keypad to
enter the correct value.
Repeat until all fields are correct then press
ENTER
19