background image

Donning the Blo

w

er/Lo

w Battery Alarm

8

Donning the Blower and Respirator

Prepare to don the blower, battery and hood in a safe, hazard-free
area and do the following:

Check that the cartridges are properly mounted on the blower
unit.

Prior to assembling the system, place the battery in the battery
compartment on the back of the blower. Make sure that the correct
battery is used in the apporpriate blower unit. One or two battery
packs may be used (See instructions and warning on page 4). Fit
the blower and belt around the user’s waist. With the blower at
the rear of the user, adjust the belt for a comfortable fit.

Remove the belt and blower.

Ensure that the cartridges used are suitable for the contaminant
in question and are compatible with the PA3BU/PA3ISBU Blower
Unit.

For Hoods:

Insert the PA1BT Breathing Tube approximately 5 inches into the
air entry sleeve at the rear of the hood being used 

(see Figure 7)

.

Install nylon clamp S18051 over air entry sleeve and breathing
tube, inserting clamp locks through two holes in plastic anchor
plate that is sewn into hood (RT Hoods DO NOT have a plastic
anchor plate.). 

Engage clamp locks and squeeze together until tight.

For Loose Fitting Facepieces:

If the loose fitting facepiece hood is being used, insert the bayonet
connector of the PA20LFBT Breathing Tube into the hood connector
and turn clockwise until it locks in place 

(see Figure 8)

.

For All Headpieces:

Attach the other end of breathing tube to blower unit by screwing
adapters together.

Remove any protective film covering the visor of the hood.

Put on the belt and blower assembly and make any final
adjustments to the belt as necessary, keeping the breathing tube
and hood behind the head.

Switch 

ON

the blower.

Place the hood on the head making any final adjustments to the fit
as required at this time to ensure a comfortable and stable fit.

PA3BU/PA3ISBU Low Voltage Alarm

Both the PA3BU and PA3ISBU Blower units are equipped with a
Low Voltage Alarm. This device will activate if the battery voltage
is below acceptable levels. The PA3BU Low Voltage Alarm is an
electronic 77 dba beep. The PA3ISBU Low Voltage Alarm is a
mechanical pulsing of the blower. In either case, the alarm is
internal to the blower and the sound will be carried up the
breathing tube into the hood. Sounding of the alarm indicates that
insufficient airflow may be imminent. The user should immediately
do the following:

Leave the hazard area, remove the headpiece, disconnect the
breathing tube from the hood and check the airflow with the
airflow indicator (see page 5).

If the airflow indicator indicates insufficient airflow, the battery
should be fully charged (see “Battery Pack” on page 4), and/or the
filter/cartridge should be replaced.

The PA3BU/PA3ISBU Low Voltage Alarm must not be solely relied
upon as an indication of a low air flow condition. Only the Air Flow
Indicator should be utilized for checking for adequate air flow, as
required by NIOSH.

NOTE

The PA3BU/PA3ISBU blower is provided with a circuit to
protect the battery. It will not allow the battery to be
discharged below a safe voltage for the cells, regardless
of airflow, without the Alarm sounding. When the Low
Voltage Alarm sounds and the filter cartridges are not
clogged, the battery should be recharged to protect the
battery and thereby prolong the working life of the unit.
If the ball in the Airflow Indicator is 

BELOW

or 

PARTLY

BELOW

the 

PASS LINE

with a fully charged battery, the

filter cartridges may need to be changed.

The use of any cartridge not approved with the
PA3BU/PA3ISBU blower units may put the user at risk and
could result in death or serious injury.

WARNING

Do not enter a hazardous area until you are sure that the blower
and hood are fully operational and the blower is running. The
user should periodically leave the hazardous area to check the
airflow through the system. If the PA3BU/PA3ISBU voltage
Alarm should sound, or if the user experiences any difficulty in
breathing, or senses any taste or any odors from the hazard, the
user should leave the hazardous area immediately. Failure to
observe these warnings could result in death or serious injury.

WARNING

www.bullard.com

Figure 7

Figure 8

 

Summary of Contents for PA30 Series

Page 1: ...espiratory protection during abrasive blasting or clean up 4 Do not use in circumstances where the airborne concentration level of contaminant exceeds maximum use concentration for this type of respirator as established by regulatory standards 5 Leave area immediately if Breathing becomes difficult Dizziness or other distress occurs You taste or smell the contaminant Unit becomes damaged Voltage a...

Page 2: ...erational Inspection 4 Mounting the Breathing Tube 5 Checking Air Flow 5 Air Purifying Elements 6 Mounting and Replacing Cartridges 7 Donning the Blower 8 Low Battery Alarm 8 Troubleshooting 9 Cleaning and Storage 9 NIOSH Approval Label 10 11 Ordering Information 12 Table of Contents 2 www bullard com ...

Page 3: ...Hat 20SICVH Hood with Taped and Sealed Seams and CBRN resistant PVC lens for use with Bullard Hard Hat 20TPC Hood with Solvent Resistant Lens and Inner Bib 20TP Hood with Solvent Resistant Lens 20LFM Loose Fitting Facepiece Hood Medium Size 20LFL Loose Fitting Facepiece Hood Large Size 6 The Battery Charger PA3SMC Quick Charger PA3GC Gang Charger PA3ISC Intrisically Safe Quick Charger PA3ISGC Intr...

Page 4: ... is fully charged This procedure should also be followed after periods of prolonged storage Without periodic charging the nickel metal hydride batteries will lose up to 1 2 of their charge per day Allowing a battery to self discharge during periods of prolonged storage will not damage the battery Pre Operational Inspection Prior to each work shift perform the following Pre Operational Inspection t...

Page 5: ... not blocked Two hands may be used if preferred one to hold the breathing tube and one to hold the Airflow Indicator The position of the ball in the Airflow Indicator should be observed If any part of the ball is below the PASS LINE on the Airflow Indicator check for Blower malfunction Clogged or damaged Air Purifying filter elements on the HE filters or the combination cartridges with HE filters ...

Page 6: ...inst the following contaminants independent laboratories have tested and verified that the PA3NBC cartridge is effective against many chemical warfare agents and or their recognized surrogates These agents surrogates are listed below WARNING Use only the cartridge described in the above table Used particulate laden cartridges must be changed as a set and not individually All cartridges must be of ...

Page 7: ...cts of the gas or vapor and the presence of any warning properties Contact Bullard for further information on change out schedules High efficiency particulate filters must be replaced when retained particles clog the filters and reduce air flow below acceptable levels as indicated by testing with the Air Flow Indicator See page 5 To Replace Cartridges Remove the air purifying element from its pack...

Page 8: ... if the battery voltage is below acceptable levels The PA3BU Low Voltage Alarm is an electronic 77 dba beep The PA3ISBU Low Voltage Alarm is a mechanical pulsing of the blower In either case the alarm is internal to the blower and the sound will be carried up the breathing tube into the hood Sounding of the alarm indicates that insufficient airflow may be imminent The user should immediately do th...

Page 9: ...ne organic based solvents or chlorinated degreasing fluids such as trichloroethylene as they will cause damage DO NOT immerse the equipment in water or other cleaning fluid as this may cause contamination in the breathing tube and blower interior that will be difficult to remove Failure to observe the instructions and warnings in this manual invalidates all performance statements and approvals for...

Page 10: ...s for changing cartridges and or filters M All approved respirators shall be selected fitted used and maintained in accordance with MSHA OSHA and other applicable regulations N Never substitute modify add or omit parts Use only exact replacement parts in the configuration as specified by the manufacturer O Refer to users instructions and or maintenance manuals for information on use and maintenanc...

Page 11: ... for changing cartridges and or filters M All approved respirators shall be selected fitted used and maintained in accordance with MSHA OSHA and other applicable regulations N Never substitute modify add or omit parts Use only exact replacement parts in the configuration as specified by the manufacturer O Refer to users instructions and or maintenance manuals for information on use and maintenance...

Page 12: ...olvent resistant polyester lens no suspension CC20 only Hood with inner bib for use with Bullard hard hat 20TICH Tychem QC hard hat not included 20SICH Tychem SL with taped and sealed seams hard hat not included 20SICVH Tychem SL with taped and sealed seams and CBRN resistant PVC lens hard hat not included Hood with inner bib and long outer bib for use with headband suspension 20TICS Tychem QC wit...

Reviews: