background image

9

PA30 Series Powered Air-Purifying Respirator

PA3/PA3IS Blower Assembly User Manual

Tr

oubleshooting/Cleaning and Stor

age

Troubleshooting

The following guide will enable you to locate and correct malfunctions:

Possible Cause

Solution

Low Voltage

Re-charge battery.

Air inlet to hood is twisted or constricted,

Remove, inspect, and reinsert PA1BT

or breathing tube inserted incorrectly

Breathing Tube approximately 5" into 
hood inlet.

Clogged/damaged air-purifying 

Replace the filter cartridge.

filter element

Battery low

Re-charge battery.

Blower malfunction

Replace blower.

Breathing tube or hood damaged

Replace breathing tube and/or hood.

Equipment damaged or filter

Leave hazardous area immediately 

needs to be replaced

and check equipment.

Replace filter. 

If the problem persists and no 
damage is found, return equipment 
for repair.

Low airflow

See above.

Improper initial charge

Review “Initial Charging Procedure” 

Prolonged storage of unit not on charger

on page 4.

Cleaning

Avoid contaminant entry into the breathing tube, as this will
compromise respiratory protection and could result in death or
serious injury. Consult your local safety professional if you
suspect that contaminant has entered the breathing tube. When
cleaning the equipment, do the following:

Ensure water does not enter cartridges. Replace wet cartridges.

DO NOT

use gasoline, organic-based solvents, or chlorinated

degreasing fluids (such as trichloroethylene) as they will cause
damage.

DO NOT

immerse the equipment in water or other cleaning fluid

as this may cause contamination in the breathing tube and
blower interior that will be difficult to remove.

Failure to observe the instructions and warnings in this manual
invalidates all performance statements and approvals for this
equipment and could result in death or serious injury.

Once cartridges have reached the end of their useful life, discard
in accordance with federal, state, and local guidelines, and  in
conformance with plant safety regulations. 

Use a lint-free cloth moistened in a mild solution of soap and warm
water to clean the outer surface of the equipment.

Consult the CC20/RT Series Tychem Hood User Manual for
cleaning instructions for the hood components. 

Storage

When the apparatus is completely dry, store in a clean, dry area,
away from direct sunlight and sources of direct heat.

The storage temperature should be between 32º F to 90º F (0º C
to 32º C) with humidity less than 90% RH.

Consult the CC20/PA30 Series Tychem Hood User Manual for
storage instructions on hood components.

WARNING

Malfunction

Low Voltage Alarm is sounding, but
Airflow Indicator shows adequate
air flow

No/low airflow into covering
(Alarm sounding for low voltage)

Smell or taste contaminant

Blower unit does not run for 
full service life
PA3 (8-10 hours with 2 battery packs;

4-5 hours with 1 battery pack)

PA3IS (Approximately 7 hours with 1

battery pack)

www.bullard.com

 

Summary of Contents for PA30 Series

Page 1: ...espiratory protection during abrasive blasting or clean up 4 Do not use in circumstances where the airborne concentration level of contaminant exceeds maximum use concentration for this type of respirator as established by regulatory standards 5 Leave area immediately if Breathing becomes difficult Dizziness or other distress occurs You taste or smell the contaminant Unit becomes damaged Voltage a...

Page 2: ...erational Inspection 4 Mounting the Breathing Tube 5 Checking Air Flow 5 Air Purifying Elements 6 Mounting and Replacing Cartridges 7 Donning the Blower 8 Low Battery Alarm 8 Troubleshooting 9 Cleaning and Storage 9 NIOSH Approval Label 10 11 Ordering Information 12 Table of Contents 2 www bullard com ...

Page 3: ...Hat 20SICVH Hood with Taped and Sealed Seams and CBRN resistant PVC lens for use with Bullard Hard Hat 20TPC Hood with Solvent Resistant Lens and Inner Bib 20TP Hood with Solvent Resistant Lens 20LFM Loose Fitting Facepiece Hood Medium Size 20LFL Loose Fitting Facepiece Hood Large Size 6 The Battery Charger PA3SMC Quick Charger PA3GC Gang Charger PA3ISC Intrisically Safe Quick Charger PA3ISGC Intr...

Page 4: ... is fully charged This procedure should also be followed after periods of prolonged storage Without periodic charging the nickel metal hydride batteries will lose up to 1 2 of their charge per day Allowing a battery to self discharge during periods of prolonged storage will not damage the battery Pre Operational Inspection Prior to each work shift perform the following Pre Operational Inspection t...

Page 5: ... not blocked Two hands may be used if preferred one to hold the breathing tube and one to hold the Airflow Indicator The position of the ball in the Airflow Indicator should be observed If any part of the ball is below the PASS LINE on the Airflow Indicator check for Blower malfunction Clogged or damaged Air Purifying filter elements on the HE filters or the combination cartridges with HE filters ...

Page 6: ...inst the following contaminants independent laboratories have tested and verified that the PA3NBC cartridge is effective against many chemical warfare agents and or their recognized surrogates These agents surrogates are listed below WARNING Use only the cartridge described in the above table Used particulate laden cartridges must be changed as a set and not individually All cartridges must be of ...

Page 7: ...cts of the gas or vapor and the presence of any warning properties Contact Bullard for further information on change out schedules High efficiency particulate filters must be replaced when retained particles clog the filters and reduce air flow below acceptable levels as indicated by testing with the Air Flow Indicator See page 5 To Replace Cartridges Remove the air purifying element from its pack...

Page 8: ... if the battery voltage is below acceptable levels The PA3BU Low Voltage Alarm is an electronic 77 dba beep The PA3ISBU Low Voltage Alarm is a mechanical pulsing of the blower In either case the alarm is internal to the blower and the sound will be carried up the breathing tube into the hood Sounding of the alarm indicates that insufficient airflow may be imminent The user should immediately do th...

Page 9: ...ne organic based solvents or chlorinated degreasing fluids such as trichloroethylene as they will cause damage DO NOT immerse the equipment in water or other cleaning fluid as this may cause contamination in the breathing tube and blower interior that will be difficult to remove Failure to observe the instructions and warnings in this manual invalidates all performance statements and approvals for...

Page 10: ...s for changing cartridges and or filters M All approved respirators shall be selected fitted used and maintained in accordance with MSHA OSHA and other applicable regulations N Never substitute modify add or omit parts Use only exact replacement parts in the configuration as specified by the manufacturer O Refer to users instructions and or maintenance manuals for information on use and maintenanc...

Page 11: ... for changing cartridges and or filters M All approved respirators shall be selected fitted used and maintained in accordance with MSHA OSHA and other applicable regulations N Never substitute modify add or omit parts Use only exact replacement parts in the configuration as specified by the manufacturer O Refer to users instructions and or maintenance manuals for information on use and maintenance...

Page 12: ...olvent resistant polyester lens no suspension CC20 only Hood with inner bib for use with Bullard hard hat 20TICH Tychem QC hard hat not included 20SICH Tychem SL with taped and sealed seams hard hat not included 20SICVH Tychem SL with taped and sealed seams and CBRN resistant PVC lens hard hat not included Hood with inner bib and long outer bib for use with headband suspension 20TICS Tychem QC wit...

Reviews: