4.5 Connecting the Gas Line
The sample gas line must be carefully and properly connected using a suitable fitting.
This table provides an overview of the sample gas probe connections:
Probe
GAS 222
Reservoir
PAV01
Ball valve
pneumatic drive
Control valve
3/2-way solenoid
valve
Connecting flange
1)
DN65/PN6/
DN3"-150
Sample gas inlet
G3/4
Sample gas outlet
NPT 1/4
Blowback connection
G3/8
Test gas connection
1)
Tube Ø6 mm
Tube Ø1/4
Filling port
NPT 1/4
Condensate
G1/2
Bypass
NPT 1/4
Control air
G1/8
G1/4
NPT 1/4
Tab. 1:
Gas Probe Connections (Varies by Model)
1)
Varies by version.
Connecting the sample gas line requires a pipe fitting with NPT 1/4" female thread.
WARNING
Gas emanation
Sample gas can be harmful to the health!
Check the lines for leaks.
4.5.1 Blowback Connection
If the probe has a G3/8 blowback connection, without factory installed blowback device it will be open. The blowback connection
must be sealed gas-tight prior to start-up.
DANGER
Toxic, corrosive gasses
Explosive or toxic gases can develop due to a leaking or open blowback connection.
4.5.2 Connecting the calibrating gas line (optional)
Connecting the calibrating gas line requires a Ø6 mm or Ø1/4” pipe fitting.
If the calibrating gas connection was ordered with check valve, a Ø6 mm or Ø1/4” pipe can be connected directly to the check
valve.
9
Bühler Technologies GmbH
BE460064 ◦ 08/2019
Summary of Contents for GAS 222.30 Ex2
Page 23: ...GAS 222 30 Ex2 9 5 Dimensions 21 B hler Technologies GmbH BE460064 08 2019...
Page 24: ...GAS 222 30 Ex2 22 B hler Technologies GmbH BE460064 08 2019...
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 44: ...DE 461099 05 2016 Page 8 9...