4.8 Analog
Outputs
4.8.1 Analog Output x: Setting the upper Limit
Here, the upper limit of the measured level is assigned to the maximum analog output signal. The
parameter is set in the following menu:
Range:
[
o.Lo
] … [
o.Hi
]
Note:
-
The set output range must not be smaller as 10% of the measuring
range:
A1.Hi
–
A1.Lo
>= 10% * (
o.Hi
–
o.Lo
)
-
If the output range is set to small, the analog output signal may show
steps.
4.8.2 Analog Output x: Setting the lower Limit
Here, the lower limit of the measured level is assigned to the minimum analog output signal. The
parameter is set in the following menu:
Range:
[
o.Lo
] … [
o.Hi
]
Note:
-
The set output range must not be smaller as 10% of the measuring
range:
A1.Hi
–
A1.Lo
>= 10% * (
o.Hi
–
o.Lo
)
-
If the output range is set to small, the analog output signal may show
steps.
4.8.3 Analog Output x: Setting the Output Characteristic
The output characteristic of the analog output may be defined as voltage output or as current output with
different signal output ranges. The output characteristic is set in the following menu:
The following options are available:
4 mA to 20 mA
2 V to 10 V
0 V to 10 V
0 V to 5 V