36
MA112780_38.6
[Original Instructions]
9/19/2011
Automatic Blade Dressing
The Automatic Blade Dressing system uses a fine-tooth shaft for advancing the Dressing Stick,
giving maximum cleaning cuts from the stick. Periodically this fine-toothed shaft will need to be
clean.
Cleaning the Automatic Blade Dressing shaft
1. Turn off the IsoMet 5000.
2. Open the hood.
3. Pull the T-mount out of the housing.
4. Using the supplied cleaning brush, clean the shaft thoroughly and lubricate with a light oil.
Do
not
use WD40™.
5. Carefully align T-mount and push it back into the housing.
Checking the Blade Motor Total Hours
To check the blade motor total operating time in hours:
POWER
OFF
ON
IsoMet 5000
LINEAR PRECISION SAW
R
Figure 13
Hidden Button on the Front Display Panel
1. There is a hidden, non-tactile button located on the front display panel.
2. Power off the IsoMet 5000.
3. Press and hold the hidden button on the front display panel, this will power on the machine.
4. The LCD Screen will now display BLADE HOURS. This is the total number of hours that the
blade motor has been rotating.
5. Press the ZERO button to return to the normal operating screens.
Hidden Button
(No Dome)