MA112680_39.3
4/2014
www.buehler.com
IsoMet™ 4000
Linear Precision Saw
Page 1: ...MA112680_39 3 4 2014 www buehler com IsoMet 4000 Linear Precision Saw ...
Page 2: ...en a leading manufacturer of scientific instruments and supplies for use in materials analysis Buehler products are used throughout the world in manufacturing facilities quality laboratories and universities to analyze all types materials including Ferrous and Non ferrous metals Thermal spray coatings Printed Circuit Boards Fasteners Ceramics Composites Semiconductors Rocks Glasses Plastics Compan...
Page 3: ...06 EN 61000 4 4 2012 EN 61000 4 5 Ed 2 0 2006 EN 61000 4 6 Ed 2 2 2009 EN 61000 4 8 2010 Quality Management System ISO 9001 2008 Registered firm Underwriters Laboratories Inc QMS Cert 10001679 1130 W Lake Cook Road Suite 340 Buffalo Grove IL 60089 USA This machine is CE marked RoHS Compliant 2011 65 EU Directive Lake Bluff Illinois USA Prepared by Kate Watling Technical Communicator THIS MANUAL IS...
Page 4: ...This page was intentionally left blank iv MA112680_39 3 Original Instructions 4 2014 ...
Page 5: ...reen Commands 12 L3 Display Screen 13 L3 Display Screen Commands 13 Specimen Positioning 14 Specimen Loading 14 Positioning a Specimen with an Unknown Thickness 14 Positioning a Specimen with a Known Thickness 16 Operation 17 Cutting a Specimen 17 Single Cut 17 Manual Cutting 18 SMART CUT Checking and Adjusting the Feed Rate 19 Blade Dressing 19 Dressing the Blade 20 Automatic Blade Dressing Rotat...
Page 6: ...rnal Coolant Lubricant Recirculating System 25 Draining the Coolant 25 Cleaning the Unit 26 Automatic Blade Dressing 26 Cleaning the Automatic Blade Dressing shaft 26 Checking the Blade Motor Total Hours 27 Troubleshooting Chart 28 Application Guide 29 Buehler Environmental Policy 31 ...
Page 7: ...ate of receipt by the customer The warranty is void if inspection shows evidence of abuse misuse unsafe use or unauthorized repair This warranty covers all Buehler costs associated with the replacement of defective materials e g parts and labor If for any reason this unit must be returned to Buehler for warranty service please contact Buehler Service at www Buehler com or 1 800 283 4537 for prior ...
Page 8: ...le material must not be used with the IsoMet 4000 Linear Precision Saw Maintain the IsoMet 4000 Linear Precision Saw with care Properly maintained machines are less likely to bind and are easier to control Any alteration or modification is a misuse and may result in a dangerous situation Maintain machine guards and interlocks Do not attempt to enter the cutting bay when the IsoMet 4000 Linear Prec...
Page 9: ...696 Buehler Recommended Cutting Fluid Dressing Stick 11 1190 IsoCut 7 inch Wafering Blade 11 4267 Set of 4 inch Stainless Steel Flanges 11 2689 Short T Slot Table 11 2701 1µm Specimen Positioning System 11 2699 2 Abrasive Cut off Wheels 11 4207 010 11 4217 010 3 Specimen Chucks 11 2683 Single Saddle Chuck 11 2684 1 inch Round Specimen Chuck 11 2686 Irregular Specimen Chuck DANGER indicates a hazar...
Page 10: ...ical Installation Electrical Shock Hazard Only a qualified electrical technician should perform electrical installation and maintenance Electrical Shock Hazard Do not change the power plug in any way Buehler machines are equipped with a polarized plug one blade is wider than the other and a ground pin Polarized plugs reduce the risk of electrical shock This plug will fit in a polarized outlet only...
Page 11: ... end cap bushing from drive shaft see Figure 1 2 Install the outer flange on the end cap bushing 3 Install the blade against the outer flange 4 Slip on the inner flange on the end cap bushing 5 Tighten the end bolt to complete the installation NOTICE To prevent misalignment and damage to the blade clean the end cap bushing screws and flanges in a mild detergent solution to remove particles from pr...
Page 12: ...water Do not over fill the coolant tank The true level of the liquid is indicated on the front of the tank and not at the drain inside the cutting chamber During high frequencies of use or when using abrasive cut off wheels with the IsoMet 4000 Linear Precision Saw the coolant tank should be cleaned regularly The drain screen should be cleaned between each cut to prevent blockage For high frequenc...
Page 13: ...outlet pipe Figure 3 Supply fitting on the recirculating tank 5 Connect the power cord from the saw to the 12 volt power plugs from the saw to the banana plugs on the recirculating tank see Figure 3 a Plug the BLACK banana plug into the BLACK connector b Plug the RED banana plug into the RED connector 6 Fill the recirculating tank with seven 7 gallons of mixed coolant or until the mixed coolant re...
Page 14: ...n The front panel LCD will light up and scroll through the display screen Raise the hood to allow access to the cutting bay Front Panel Controls POWER OFF ON R IsoMet 4000 LINEAR PRECISION SAW 1 2 3 4 5 6 EMERGENCY STOP The front panel consists of six 6 dedicated buttons five 5 software multi function softkey buttons an Emergency Stop button and a large Liquid Crystal Display LCD with back lightin...
Page 15: ...will change to 00 indicating the SOFT HOME position Increase The Increase button will incrementally increase raise a parameter s value The maximum values for the IsoMet 4000 are BLADE SPEED 5000 rpm FEED RATE 75 in min 19 mm CUTTING DISTANCE 8 00 inches 203 mm Decrease The Decrease button will incrementally decrease lower a parameter s value The minimum values for the IsoMet 4000 are BLADE SPEED 2...
Page 16: ...s BLADE SPEED 4250rpm FEED RATE 64in min DISTANCE REMAINING 1 00in CUTTING CYCLE ON BLADE MOTOR ON PUMP MOTOR ON FEED MOTOR ON L1 BLADE SPEED FEED RATE and DISTANCE REMAINING parameter fields BLADE SPEED FEED RATE CUTTING LENGTH Press the SCROLL button to display the CUTTING LENGTH parameter Softkey Buttons activate or deactivate a function Parameter fields are highlighted using the Scroll button ...
Page 17: ... ON or deactivates OFF the coolant pump motor This is the only function that will work while the hood is open Button 4 FEED MOTOR activates ON or deactivates OFF the feed motor The FEED MOTOR can be used to position the blade and will only operate when the blade is rotating Button 5 Scrolls the LCD Screen between Screen L1 Screen L2 and Screen L3 Pause CUTTING CYCLE To pause the cutting cycle pres...
Page 18: ...tops and pauses the cutting cycle Button 2 ROTATING CHUCK activates ON or deactivates OFF The chuck will rotate at a constant speed Button 3 SOFT START selects the Soft Start option This will slow the selected FEED RATE for the first 06 inch 1 52 mm of a cut Button 4 SOFT STOP selects the Soft Stop option This will slow the selected FEED RATE for the last 25 inch 6 35 mm of a cut Button 5 Scrolls ...
Page 19: ...pauses the cutting cycle Button 2 DRESS BLADE sets the correct blade dressing parameters and activates the automatic dressing mechanism if attached Button 3 UNITS changes between METRIC and IMPERIAL English units of measure Button 4 LANGUAGE allows the operator to select a language from the list The languages currently available are English French German Spanish Portuguese Japanese Chinese Korean ...
Page 20: ... 2 Determine the cut length of the specimen a Position the specimen in the chuck so the cutting blade can pass by the chuck without touching the specimen see Figure 5 b Turn the hand crank counter clockwise to bring the blade forward until it is just behind the and almost touching the specimen see Figure 6 c Press the ZERO button on the control panel and the DISTANCE REMAINING field value will dis...
Page 21: ...ighlighted b Use the INCREASE or DECREASE buttons to change the value to equal the REMAINING DISTANCE value The value will automatically be saved 4 Turn the crank clockwise to position the blade behind the specimen The value under REMAINING DISTANCE does not have to equal 00 in or 00 mm after repositioning the blade 5 Reposition the specimen and tighten the vise to prevent the specimen from moving...
Page 22: ...t the specimen from moving 5 Press the ZERO button on the control panel and the DISTANCE REMAINING field value will display 00 in or 00 mm 6 Program the known distance into the CUTTING LENGTH field a Press the SCROLL button until the CUTTING LENGTH parameter field is highlighted b Use the INCREASE or DECREASE buttons to change the value to equal the REMAINING DISTANCE value The value will automati...
Page 23: ...ING LENGTH Press the SCROLL button until the desired parameter is highlighted and adjust the parameter values using the INCREASE or DECREASE buttons 2 Position the specimen using any of the required chucks 3 Use the SCROLL button to scroll to the L1 Screen 4 Press the PUMP MOTOR button to the ON position and position the coolant flow as desired 5 Press the CUTTING CYCLE button to the ON position T...
Page 24: ...Cutting There are two 2 ways to manually cut a specimen Method 1 1 Mount the specimen as previously described 2 Press the BLADE MOTOR and PUMP MOTOR buttons to the ON position 3 Adjust the BLADE SPEED to the desired rpms 4 Manually turn the crank to cut the specimen 5 Retract the blade when the cut is complete 6 Press the BLADE MOTOR and PUMP MOTOR buttons to the OFF position Method 2 1 Mount the ...
Page 25: ...ing exposes the abrasive grain to ensure free cutting New wafering blades should be dressed several times and older blades dressed as required based on the specimen material properties The IsoMet 4000 is shipped with the Automatic Blade Dressing attachment adjusted for a 7 inch diameter blade When a different diameter blade is installed the Automatic Blade Dressing attachment must be readjusted Fi...
Page 26: ...nto the dressing stick Continue to cut by advancing the blade forward into the piece and press the FEED MOTOR button to the ON position If the Automatic Dressing System is attached and a dressing stick is present advance to the L3 Screen and press the DRESS BLADE button If not cutting this will automatically run the blade at the proper rpm and activate the lubricant The Automatic Dressing System w...
Page 27: ... and a Rotating Chuck There are three 3 power sockets in a vertical line located behind the control housing The top power socket is not used in the IsoMet 4000 The middle power socket connects to the Rotating Chuck The bottom power socket connects to the Automatic Blade Dressing Figure 12 IsoMet Power Sockets Removing the Specimen Position System Catalog Number 11 2699 1 Turn OFF the IsoMet 4000 2...
Page 28: ...h provided until tight Removing the Rotating Chuck Catalog Number 11 2695 1 Turn OFF the IsoMet 4000 2 Use an Allen wrench 4 metric to loosen the two 2 mounting screws 3 Disconnect the plug from the control panel 4 Recap the middle power socket Installing the Rotating Chuck 1 Insert the mounting T nuts into the T slot of the rail parallel to the turret 2 Set the Rotating Chuck assembly on to the T...
Page 29: ...materials and fluids 2 Place the Automatic Blade Dressing unit with the mounting flange down and to the right 3 Align the mounting holes 4 Insert the screws and tighten 5 Uncap the power socket receptacle 6 Insert the power plug and secure it 7 Attach a dressing stick using the thumbscrew 8 Adjust the T mount to the blade size being used ...
Page 30: ...lade s forward travel Use the hand crank to reposition the blade BLADE PINCHED Indicates that during cutting the blade became pinched by the specimen and the cutting operation has automatically shut off Carefully retract the blade and check the specimen fixture EMERGENCY STOP Indicates the Emergency Stop button has been pressed 1 Examine the machine for problems 2 Rotate the Emergency Stop button ...
Page 31: ...Follow the mixing directions for the Buehler recommended cutting fluid as indicated on the cutting fluid container Draining the Coolant While the machine is not in use remove the coolant tank remove the lid and pour the coolant in an appropriate container 1 Lift the hood 2 Pull off the Coolant Tube fitting and hold it over a container 3 Press the PUMP MOTOR button to the ON position 4 Let the pump...
Page 32: ...th non abrasive household cleaners The protective hood and touch pad control panel should be cleaned using mild soap and water applied with a soft cloth Automatic Blade Dressing The Automatic Blade Dressing system uses a fine tooth shaft for advancing the Dressing Stick giving maximum cleaning cuts from the stick Periodically this fine toothed shaft will need to be cleaned and lubricated Cleaning ...
Page 33: ...y Panel 1 There is a hidden non tactile button located on the front display panel 2 Power off the IsoMet 4000 3 Press and hold the hidden button on the front display panel this will power on the machine 4 The LCD Screen will now display BLADE HOURS This is the total number of hours that the blade motor has been rotating either cutting or not 5 Press the ZERO button to return to the normal operatin...
Page 34: ...CUTTING DISTANCE The DISTANCE REMAINING may be 00 or less Increase the DISTANCE REMAINING The ARM LIMIT switch may be activated Deactivate the ARM LIMIT switch The bed has rust on it Improper coolant mixture Check the coolant for proper mixture The hood has been closed too long with moisture build up Open hood and keep hood open when not in use The Specimen Positioning System does not move Cable i...
Page 35: ...500 1 00 0 50 13 mm Partially stabilized zirconia ZrO2 Series 15 LC Low 2500 1 15 0 40 10 mm Printed circuit boards 11 4217 Abrasive Wheel 4000 1 15 0 40 10 mm Sapphire Al2O3 Series 15 LC Low 1500 1 15 0 40 10 mm Silicon carbide SiC Series 15 LC Low 2500 1 00 0 50 13 mm Thermal spray coatings 15 HC or 11 4207 Abrasive Wheel High 3000 4000 2 00 1 15 0 25 6 mm 0 40 10 mm 5 Titanium alloy Series 15 H...
Page 36: ...40 10 mm Turbine blades Titanium 114 217 Abrasive wheel 4000 1 15 0 40 10 mm Zinc 11 4217 Abrasive wheel 4000 1 15 0 40 10 mm 0 29 6 mm x 125 3 mm rectangular section 2 0 67 17 mm diameter 3 0 570 15 mm diameter 4 0 74 19 mm diameter 5 1 25 mm x 25 6 mm rectangular section 6 0 375 0 95 mm diameter 7 BU01 BU24 based on a 5 inch blade 8 BU25 BU35 based on a 7 inch blade Reduce speed to 0 2 min 5 mm ...
Page 37: ...ocesses our materials and our people in order to reduce the environmental impacts associated with our products To help conserve natural resources and to protect human health and environment please follow your state and local regulations on recycling and disposing of waste consumables or parts related to your Buehler machine For End Of Life on Buehler machines if recycling and disposal facilities a...
Page 38: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 32 MA112680_39 3 Original Instructions 4 2014 Figure 14 Control Housing Sub Assembly Diagram 2680S803 ...
Page 39: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 33 MA112680_39 3 Original Instructions 4 2014 Figure 15 Hood Sub Assembly Diagram 2680S802 ...
Page 40: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 34 MA112680_39 3 Original Instructions 4 2014 Figure 16 Turret Sub Assembly Diagram 2680S800 ...
Page 41: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 35 MA112680_39 3 Original Instructions 4 2014 ...
Page 42: ...wings Note Drawings and Parts List are subject to change without notice 36 MA112680_39 3 Original Instructions 4 2014 Figure 17 Spindle Sub Assembly Diagram 2680S806 Figure 18 Turret Core Sub Assembly Diagram 2780S805 ...
Page 43: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 37 MA112680_39 3 Original Instructions 4 2014 Figure 19A Feed Screw Sub Assembly Diagram 2680S801 Part 1 ...
Page 44: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 38 MA112680_39 3 Original Instructions 4 2014 Figure 19B Feed Screw Sub Assembly Diagram 2680S801 Part 2 ...
Page 45: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 39 MA112680_39 3 Original Instructions 4 2014 Figure 20 Exploded View of Pump Assembly Diagram 2780S804 ...
Page 46: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 40 MA112680_39 3 Original Instructions 4 2014 Figure 21 IsoMet 4000 Top View of Assembly Diagram 2680900B ...
Page 47: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 41 MA112680_39 3 Original Instructions 4 2014 Figure 22 Bottom View of IsoMet 4000 Assembly Diagram 2680900C ...
Page 48: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 42 MA112680_39 3 Original Instructions 4 2014 ...
Page 49: ...s List are subject to change without notice 43 MA112680_39 3 Original Instructions 4 2014 Figure 23 IsoMet 4000 1um Sample Positioning System Assembly Diagram 11 2699A Figure 24 IsoMet 4000 1um Sample Positioning System Assembly Diagram 11 2699B ...
Page 50: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 44 MA112680_39 3 Original Instructions 4 2014 Figure 25 IsoMet 4000 Electrical Diagram 2680900D ...
Page 51: ...IsoMet 4000 Drawings Note Drawings and Parts List are subject to change without notice 45 MA112680_39 3 Original Instructions 4 2014 Figure 26 IsoMet 4000 Packaging Diagram 2680900E ...
Page 52: ... BAG POLY TRASH 1 EA R7119 SHAKE PROOF LOCK WASHER 3 EA R9008A CORD IEC POWER U S 0 EA R9008B CORD IEC POWER EUROPE 0 EA R9008C CORD IEC POWER U K 0 EA R9008D CORD IEC POWER JAPAN 0 EA 112683 SINGLE SADDLE CHUCK STN STL 1 EA IS112682 INSTRUCTIONS 112682_83_84_85_86 1 EA L112683 LABEL FOR 112683 1 EA R8095 LABEL A SIZE 1 EA R10002 SCREW M5 X 20 SOC SS 2 EA R10474 SCREW M5 X 30 SOC HD CAP SS 2 EA R6...
Page 53: ...ULAR SPECIMEN 1 EA 112689 FLANGE SET 4 SET OF 2 SS 1 EA L112689 LABEL FOR 112689 1 EA R8095 LABEL A SIZE 1 EA R6318 POLY BAG 6X8 4 MIL 2 EA 2680S163 FLANGE 4 INCH 2 EA 112696 AUTOMATIC DRESSING SYSTEM 1 EA L112696 LABEL FOR 112696 1 EA R8095 LABEL A SIZE 1 EA R0585 TIE STRAP 10X4IN 1 EA R10097 SWITCH LIMIT 1A 12VAC 1 EA R10123 SCREW SET M5 X 12 SS 1 EA R10140 MOTOR GEAR 19VDC 41 3RPM 1 EA R10219 S...
Page 54: ...ING SYSTEM 1 EA 2680S280 MOUNT DRESSING STONE 1 EA 2680S281 NUT DRESSING SYSTEM 1 EA 2680S282 SCREW DRESSING SYSTEM 1 EA 2680S378 MACHINED DRESSING 1 EA 2680S376 CASTING DRESSING 1 EA 2680S520 CABLE BLADE DRESSING 1 EA 821545 WASHER M8 4ID 2 EA 112699 PRECISION 1UM POSITIONING SYSTEM 1 EA IS112699 INSTRUCTIONS SAMPLE POSITIONING SYS 1 EA L112699 LABEL FOR 112699 1 EA R8095 LABEL A SIZE 1 EA R0965 ...
Page 55: ...2680S365 HANDLE ISOMET 4K MICROMETER 1 EA 2680S366 TUBE ISOMET 4K MICROMETER 1 EA 114267 ISOCUT WAFERING BLADE 7X 025IN 1 EA 2150S150 RATING PLATE SMALL SELF ADHE 1 EA 2680S911 ISOMET 4000 SAW BASE UNIT ASSEMBLY 1 EA AK 207 ACCESSORY KIT F 112680 112780 1 EA LAK 207 LABEL FOR AK 207 1 EA R8095 LABEL A SIZE 1 EA MA114207 SHEET INSTRUCTION 7 C O WHEELS 1 EA R6743 CARTON BOOK FOLD 1 EA R6744 PAD 4 EA...
Page 56: ...4 EA R0615W WASHER 1 4IN SS 8 EA R0617W WASHER 5 16IN 2 EA R0618J NUT 5 16 18 HEX JAM SS 2 EA R10008 SCREW M5 X 10 SOC BUT SS 2 EA R10059 NUT M5 X 0 8 KEPS STEEL ZINC 4 EA R10237 SCREW M5 X 12 PHIL PAN HD SS 10 EA R10239 SCREW M6 X 20 PHIL PAN HD SS 4 EA R10305 CAP BACKWALL ISOMET 4000 1 EA R12063 SCREW M6 X 16 PAN HD PHIL SS 4 EA R4559 ADHESIVE ANAEROBIC THREAD GR 0 01 EA R9226 TAPE FOAM 1 8 X 3 ...
Page 57: ...BLY 1 EA R0535A TERMINAL 6 BLK SPADE 22 16 FI 2 EA R0542A TERMINAL 1 4 RING 16 14 FIG 2 EA R0605LWE WASHER EXT 6 STN STL 2 EA R0612LWE WASHER 10 EXTERNAL LOCK SS 19 EA R10079 SCREW M5 X 12 FL SOC SS 7 EA 2780S533 POWER SUPPLY 12VDC 110W 1 EA R10097 SWITCH LIMIT 1A 12VAC 1 EA R10105 CLAMP HOSE 3 EA R10106 SPRING GAS 2 EA R10107 SOCKET BALL 10MM BALL 2 EA R10108 CLIP SAFETY 10MM BALL 2 EA R10110 EYE...
Page 58: ...HESIVE ANAEROBIC THREAD GR 0 01 EA R7075 STAKON WIRE JOINTS 1 EA R7312 THERMAL COMPOUND 0 01 LB R7342 SHRINK TUBING 3 32 20 IN R7491 SCREW 6 32 X 3 8 PAN HD PHIL SS 2 EA R7760 TUBING HEAT SHRINK 25 DIA 21 IN R7845 SCREW 6 32 X 3 16 PN PHIL HD SS 4 EA R8025 TUBING HEAT SHRINK 19 DIA 20 IN R8078 SCREW 10 32 X 7 8 PAN PHIL SS 4 EA R8370 CABLE TIE MOUNT 20 EA R8814 CIRCUIT BREAKER 10A 250V 1 EA R8936 ...
Page 59: ...444 WIRE 16 GREEN YELLOW STRANDED 31 IN R0539 TERMINAL 10 RING 16 14 NIT 2 EA R0539B TERMINAL 6 RING 16 14 NIT 1 EA R0606LW LOCK WASHER 6 SS 1 EA R0612LWE WASHER 10 EXTERNAL LOCK SS 2 EA R0612W WASHER 10 FLAT SS 10 EA R10001 SCREW M5 X 10 SOC SS 25 EA R10217 BELT MICRO V 1 EA R10237 SCREW M5 X 12 PHIL PAN HD SS 2 EA R10240 SCREW M5 X 16 SOC HD SS 4 EA R4559 ADHESIVE ANAEROBIC THREAD GR 0 01 EA R11...
Page 60: ...0S805 CORE TURRET SUB ASSEMBLY 1 EA B704005 SCREW M3 X 6 REC PAN HD SS 3 EA R0609LW WASHER SPLK 8 STN STL 3 EA R10007 SCREW M4 X 10 HEX BUT HD 3 EA R10059 NUT M5 X 0 8 KEPS STEEL ZINC 4 EA R10114 SCREW M3 X 6 PAN HD PHIL 1 EA R10122 SCREW SET M5 X 6 SS 2 EA R10236 SCREW M6 X 16 SOC HEAD SS 4 EA R10381 SCREW SET M2 X 4 CUP POINT SS 2 EA R10382 SENSOR TURRET SPEED 1 EA R9408 SENSOR SPEED 1 EA R6394 ...
Page 61: ...6 LCD LOGIC CNTRL ISO4K 5K 1 EA PCB140 LCD LOGIC CNTRL ISO4K 5K 1 EA 2780P088 HEAT SINK ISOMET 4K 5K 1 EA 2780S508 CABLE 30PIN LOGIC TO MTR CNTRL 1 EA 2780S509 CABLE X1 LOGIC TO HOOD SWITCH 1 EA 2780S510 CONNECTOR MOTOR AC X1 1 EA 2780S511 CABLE AC INPUT 1 EA 2780S513 CABLE TURRET TRVL LMIT SWITCH 1 EA 2780S514 CABLE MTR CNTRL PCB TO ENCODR 1 EA 2780S517 CABLE MTR CTRL PCB PWR CNTRL 1 EA 2780S527 ...
Page 62: ...IsoMet 4000 Accessory Drawings Note Drawings and Parts List are subject to change without notice 56 MA112680_39 3 Original Instructions 4 2014 Figure 27 Automatic Dressing System Diagram ...
Page 63: ...IsoMet 4000 Accessory Drawings Note Drawings and Parts List are subject to change without notice 57 MA112680_39 3 Original Instructions 4 2014 Figure 28 Automatic Dressing System Diagram Detail A ...
Page 64: ...IsoMet 4000 Accessory Drawings Note Drawings and Parts List are subject to change without notice 58 MA112680_39 3 Original Instructions 4 2014 Figure 29 Dressing System Mounting Hole Diagram ...
Page 65: ...uehler Canada 10 Carlow Court 2 Whitby Ontario L1N 9T7 Tel 800 268 3593 905 430 4684 Fax 888 268 0371 905 430 4647 Buehler Canada Website www buehler ca Email info buehler ca Service service buehler ca Buehler France Tel 33 0 800 89 73 71 Fax 33 0 800 88 05 27 Buehler France Website www buehler fr E mail info fr buehler com Buehler France Service 69570 Dardilly France Tel 04 37 59 81 20 Fax 04 37 ...
Page 66: ...Notes 60 MA112680_39 3 Original Instructions 4 2014 ...