Buckleys PST-100 Instruction Manual Download Page 7

PST-100 

In a reasonably ventilated room of a volume greater than 40m

3

, the exposure to 

ozone produced by the high voltage spark should not present a risk to health. In a 

‘confined space’ situation it is likely that ozone levels will exceed the exposure limit 

and present some risk to health. Under these conditions adequate ventilation must 

be provided. 

 

 

 

 

Misuse or failure to comply with the guidelines outlined in this manual may impair 

the safety provided by the equipment. 

PST-100 Earth Terminal 

The PST-100 has been tested and verified by an authorised, independent third 

party as meeting the requirements of IEC 61010:2010 

- Safety requirements for 

electrical equipment for measurement, control, and laboratory use. 
The construction methods employed throughout comply with double insulation. 

The earth terminal located on the handle is a functional earth terminal, NOT a 

Protective Conductor Terminal as defined by IEC61010 and is intended as a high 

voltage return point during the operation of the PST-100. This arrangement also aids 

suppr

ession of possible EMC interference.

 

Therefore when conducting an electrical safety test (Portable Appliance Test) to 

the PST-100 it should be considered a class 2 appliance and tested accordingly. 
 
 

WARNING: 

This equipment must not be used in any combustible 

atmosphere; the high voltage discharge will cause a spark from which 

an explosion could result. 

WARNING: 

This equipment must not be exposed to damp or wet conditions, or 

where the amount of conductive dust is greater than would occur in a normal 

situation 

WARNING: 

Never connect or disconnect the electrode with the PST-100 

connected to the mains supply. 

IMPORTANT: 

We strongly advise that individuals with pacemakers, cochlear 

implants or trans-dermal electronic implants - e.g. glucose monitors do not use 

our high-voltage test equipment under any circumstances. 

Summary of Contents for PST-100

Page 1: ...PST 100 High Voltage Spark Tester Instruction manual Please read this manual before using the equipment 5830 0012 07 21 English...

Page 2: ......

Page 3: ...ating the test voltage 9 SP 02 74 test voltage formula 9 Operation 10 Setting the Output Voltage 10 Applications 11 Insulation Testing 11 Gas Discharge Lamp Testing 11 Ozone Production 11 Risk Assessm...

Page 4: ...or flaws in electrically non conductive coatings linings membranes pipe wraps and other plastic coated metals The PST 100 generates a pulsed ring output voltage The pulse repetition is 20mS 50Hz or 16...

Page 5: ...0007 6000 0015 105 120V AC Supply frequency 50 to 60 Hz Max power consumption 30VA Output voltage 10kV to 55kV Output Pulsed Ring Pistol weight 1 0kg Altitude Up to 2000m Relative humidity 80 Max Non...

Page 6: ...sed by responsible and authorised personnel that have read and understood this manual The probe must NEVER be directed at the body There is a potential risk to those who might have an incipient heart...

Page 7: ...handle is a functional earth terminal NOT a Protective Conductor Terminal as defined by IEC61010 and is intended as a high voltage return point during the operation of the PST 100 This arrangement als...

Page 8: ...to a point of a measuring or control circuit or to a screening part and which is intended to be earthed for any functional purpose other than safety Double Insulation Insulation comprising both basic...

Page 9: ...spark then forms ionising the air drastically reducing its electrical resistance For materials with a thickness in the range of 1mm to 30mm the NACE Standard SP 02 74 voltage calculation formula is no...

Page 10: ...be kept moving when testing as the high voltage and ring frequency combination may cause heating in the article under test which may result in burning The output control should be adjusted for the lo...

Page 11: ...probe discharge will change to a single spark Remove the probe from the surface and mark the point where the fault was detected Gas Discharge Lamp Testing The PST 100 can be used to excite gas dischar...

Page 12: ...rks is the test outdoors or is there sufficient suitable ventilation in place to ensure that this is not hazardous Is the access route to the area of the test safe Is fall protection necessary If so i...

Page 13: ...the rear panel These may be replaced by the user Disconnect the PST 100 from the mains supply before attempting to replace fuses Disposal information Producer registration number WEE HJ0051TQ This Pr...

Page 14: ...CEO Buckleys UVRAL Ltd UKCA Declaration of conformity We Buckleys UVRAL Ltd as sole manufacturer of the apparatus listed below declare that the product PST 100 High Voltage Spark Tester is manufactur...

Page 15: ...nufacturer Buckleys UVRAL Ltd Address Buckleys House Unit G Concept Court Shearway Business Park Shearway Road Folkestone Kent CT19 4RG UK Tel 44 0 1303 278888 Website www buckleysinternational com Di...

Page 16: ...ng to your product Be the first to hear about our new products We strive to improve the quality of our products and service Registering your product helps us monitor overall quality of our products se...

Reviews: