3 Technical data
14
V-850 / V-855 Operation Manual, Version E
3 .2
Technical data overview
Technical data
Vacuum Controller V-850 / V-855
Dimensions casing ( W×H×D)
160×115×120 mm
Weight
700 g
Power supply
30 VDC from RJ-45 connector when using Controller
with a Rotavapor R-210 / R-215 or Vacuum Pump V-700
/ V-710 or from the power pack for stand-alone versions.
Frequency
50/60 Hz
Power consumption
10 W
Interfaces
USB (data transmission), RS 232 / RS 485 (communica-
tion), remote control, cooling water valve, Switch Box,
vacuum valve
Environmental conditions
For indoor use only 10-40 °C, up to 2000 m, maximum
relative humidity 80 % for temperatures up to 31 °C, and
then linearly decreasing to 50 % at 40 °C
Magnetic valve supply
24 V
Vacuum connection
GL14
Measuring principle
Capacitive
Measuring range
1400–0 mbar
Control range
1100–1 mbar
Accuracy
±2 mbar (±1 digit) following alignment at constant
temperature
Temperature response
0.07 mbar/K
-1
Display range
0–1400 mbar
Hysteresis
Automatic or 1–500 mbar
Installation category
II
Pollution degree
2
Conformity
CSA/CE
Noise level
< 70 dB
3 .3
Materials used
Materials used
Component
Material designation
Pressure foil
Polyester
Casing
PBT
Diaphragm (aeration valve)
FFKM
Aeration valve
PEEK