8 Troubleshooting
52
R-210/215
Operation Manual, Version B
Table 8-2: Malfunctions with vacuum controller and vacuum pump and their remedy
Malfunction
Possible cause
Corrective measure
Frequent switching of valve or pump
System is leaky
Check all sealing points (tubes and
their connections)
Choosen hysteresis is too small
Choose larger hysteresis (if end-
vacuum is higher than 700 mbar,
switch to automatic hysteresis)
Valve does not switch
Valve does not shut off
Valve is dirty or the valve cable is not
connected
Vacuum is not reached
Back evaporation at Rotavapor
Empty receiving fl ask
Water pressure of water jet pump is
too low
Open water tap completely
Distillation “died out”
Manually decrease the pressure until
the distillation starts again
Distillation stopped, although not
dried out completely
Back evaporation from the receiving
fl ask is too strong (especially for
solvent mixtures)
Empty receiving fl ask and restart
distillation
Malfunction in distillation procedure
which is not exactly defi ned (e.g.
sudden cooling, heat fl ow too low,
etc.)
Decrease the pressure manually until
the distillation starts again
Table 8-3:
Error messages heating bath
Error number
Displayed at
Possible cause
Remedy
E01
Heating bath display
Bath temperature sensor is defective
Contact the Buchi customer service
E02
Heating bath display
Excess temperature, bath is not full
enough or empty
Switch off the instrument and let it
cool down. Fill the heating bath and
switch on the instrument again.
E03
Error is read out via the
RS-485 interface and
can be read out at the
vacuum controller
The EEPROM of the bath contains
inconsistent data or is defective
Contact the Buchi customer service
E04
Heating bath display
The bath is empty or has been fi lled
with hot liquid
Switch off the instrument and let
it cool down. Fill the bath with
heating medium and switch the
instrument on again.
E70
Heating bath display
Program error due to a software or
an electronical error
Switch the instrument off and on
again. When the error still occurs,
contact the Buchi customer
service.