![Buchi R-210 Operation Manual Download Page 38](http://html1.mh-extra.com/html/buchi/r-210/r-210_operation-manual_2825900038.webp)
5 Putting
into
operation
38
R-210/215
Operation Manual, Version B
5.8.2 Cooling
water
tubes
When connecting the white cooling water tubes,
consider the following:
Use tube clips GL-14.
The tubes used must all have the same inner
diameter (approximately 6 mm).
For safety reasons, secure the tubes with
commercial tube pivoting clamps or cable
binders.
To save cooling water, you can use the Distil-
lation Chiller B-741.
Check the tubes from time to time and
replace them when they are brittle.
•
•
•
•
•
Fig. 5.9: Cooling water tubes
5.8.3 Vacuum
tubes
When connecting the vacuum tubes, proceed
as follows:
Use tube clips GL-14.
The tubes used must all have the same inner
diameter (approximately 5 mm).
Keep vacuum tubes as short as possible.
When you are operating with the new
Buchi Vacuum Controller V-850/855 and
Vacuum Pump V-700/710 connect a Woulff
bottle between the vacuum source and the
Rotavapor.
When you are operating with a pump other
than a V-700/710, connect a valve unit to the
V-850/855 to control the vacuum.
Tubes do not need to be secured.
Check the tubes from time to time and
replace them when they are brittle.
•
•
•
•
•
•
•
Fig. 5.10: Vacuum tubes
Summary of Contents for R-210
Page 1: ...93076en R 210 215 Rotavapor Operation Manual...
Page 2: ......
Page 78: ......
Page 79: ......