![Buchi R-210 Operation Manual Download Page 50](http://html1.mh-extra.com/html/buchi/r-210/r-210_operation-manual_2825900050.webp)
7 Maintenance
and
repairs
50
R-210/215
Operation Manual, Version B
7.3.1 Cleaning
the
seals
To prolong the lifetime of the seals, rinse them regularly with water, especially if “bumping” occurred
during the distillation or if working with crystalline products. Afterwards, dry them with a solft cloth. To
remove the seals, see chapter 5.5, Installing the condenser and the seal.
7.3.2 Replacing
the
seals
Seals are subject to wear and tear, thus you should check them regularly and replace them, if neces-
sary, e.g. if they do not pass the vacuum tightness test described in chapter 5.11 anymore. For this
purpose, see chapter 5.5, Installing the condenser and the seal.
7.4 Heating
bath
The inner surface of the heating bath should be cleaned if:
The water bath is calcifi ed or contaminated.
The oil in the oil bath has changed (color, viscosity, etc.).
For this purpose, remove the heating bath from the Rotavapor and empty it.
In the case of minor calcifi cations, use a non-abrasive cleaning agent (e.g. a bathroom cleaner). If the
calcifi cation is persistent, use e.g. acetic acid to remove it. Rinse the bath thoroughly afterwards.
7.5 Glass
components
To prolong the lifetime of the glass components, consider the following:
Rinse glass components with water and commercial cleaning agent (e.g. a mild soap solution).
NOTE
We recommend cleaning all the glass components manually.
Use an alkaline cleaner to remove dirt, e.g. algae, adhering within the condenser coil.
NOTE
When a thin copper wire is introduced into the condenser coil, the risk of dirt adhering to the
condenser coil is reduced.
Remove grease from joints. After you have cleaned and completely dried each glass component,
visually inspect the components for glass splinters or cracks. Since these components are under
vacuum when the Rotavapor is operating, they are subject to strain.
Regularly check the glass components for damages and only use glassware that is in perfect condi-
tion. Glassware with cracks, stars or other damages can break during operation.
•
•
•
•
•