BTL VANQUISH ME - OPERATOR’S MANUAL I PAGE 37 OF 44
8 TECHNICAL PARAMETERS
8.1 DEVICE PARAMETERS
Name
BTL Vanquish ME
Operating conditions
Ambient temperature
+10 °C to +30 °C
(+50 °F to +86 °F)
Relative humidity
30 % to 75 %
Atmospheric pressure
700 hPa to 1 060 hPa
Position
on castors
Mode of operation
continuous
Transport and storage conditions
Ambient temperature
-10 °C to +55 °C
(+ 14 °F to + 131 °F)
Relative humidity
10 % to 85 %
Atmospheric pressure
650 hPa to 1 100 hPa
Position
vertical
Time of storage
max. 1 year
Additional conditions
transport only in original containers
Power supply
Maximum power input
600 VA
Mains voltage
~ 100 V to 120 V / ~ 200 V to 240 V
Frequency
50 Hz to 60 Hz
Electrical protection class
Class I
External exchangeable fuse
2 x T6.3 AL / 250 V, tube safety fuse 5 x 20 mm
(for supply voltage ~ 100 V to 120 V)
2 x T3.15 AL / 250 V, tube safety fuse 5 x 20 mm
(for supply voltage ~ 200 V to 240 V)
Power switch
On the back of device, positions O (off) and I (on).
To disconnect from the mains, unplug the male plug of the
power cord from the mains socket outlet.
Weight
38 kg (84 lb)
Dimensions (w x h x d)
560 x 980 x 560 mm
(22 x 39 x 22 in)
International protection rating
IP 20
Display elements
Graphic color touch screen
8.4" / 21.3 cm, 640 x 480 pixels
Indicator lights
1x orange, 4x blue
Classification
Electrical Type (Degree of Protection)
BF
FDA class
II
The manual is intended for devices with firmware version “Vanquish-MultiFirmware-v1080000”