BTL VANQUISH ME - OPERATOR’S MANUAL I PAGE 12 OF 44
Caution
With the exception of surface cleaning of the console and components and the replacement of
the fuse, the BTL Vanquish ME system has no user-serviceable parts or assemblies. Do not
remove the instrument covers under any circumstances. Call your distributor for advice about
any malfunction.
The BTL Vanquish ME is equipped with a protection system that prevents the connection of
accessories other than those supplied by the manufacturer.
The BTL Vanquish ME system does not use any drugs, creams, gels or other substances which
are applied during its use.
3.1.2
CLEANING OF THE SYSTEM CONSOLE
Switch the unit off and unplug the power plug from the power outlet. Use a soft cloth slightly
moistened with water or with a 2 % detergent solution to clean the exterior of the
BTL Vanquish ME device and its parts. Never use cleaning agents containing alcohol, ammonia,
benzine, thinners, etc. Never use abrasive cleaning materials, which will scratch the device's
surfaces.
•
In order to prevent electrical shock, unplug the power plug from the power outlet before
cleaning or disinfecting the unit.
•
Never spray aerosol cleaners directly onto the equipment. Aerosol spray cleaners are
permitted to clean the cabinet, but the cleaner should be sprayed onto a clean cloth,
which may then be used to wipe the dirty area.
3.1.3
CLEANING AND DISINFECTION OF THE ACCESSSORIES
After each patient use, clean the accessories using a soft clean cloth dampened with water and
a commercially available surface disinfectant, or use a pre-moistened wipe such as “Sani-Cloth
HB”. Avoid the use of abrasive materials, strong organic acids, oxygen-splitting compounds and
cleaning solvents.
•
Follow your local rules and regulations for appropriate disinfectant choice.
•
Wait until the console and accessories are completely dry before operating it again.