BTL – 07p
operating manual & user's guide
page 26
List of program numbers for combined ultrasound therapy/electrotherapy
51. Lumbalgie
Emitter head: 4 cm
2
. Frequency: 3 MHz, 1:4 PPR. Intensity: 0.5 W/cm
2
.
Electrotherapy: TENS cont., f = 100 Hz constant (program 1625).
Contralateral electrode placement.
Intensity above threshold sensitive.
Application time: 1 - 5 minutes, step 1 minute, daily. Total number of treatments: 3 - 5.
52. Myalgia mm. superficiales
Emitter head: 1 cm
2
. Frequency = 3 MHz, 1:2, 1:4 PPR. Intensity: 0.5 W/cm
2
.
Electrotherapy: DF, base 1 mA (program 0113).
Contralateral electrode placement.
Intensity above threshold sensitive.
Application time: 1 - 2 minutes, max. 6 minutes, daily. Total number of treatments: 2 - 3.
53. Myalgia mm. profundi
Emitter head: 1 cm
2
. Frequency = 1 MHz, 1:2, 1:4 PPR. Intensity: 1.5 - 1.7 W/cm
2
, step 0.1 W/cm
2
.
Electrotherapy: DF, base 1 mA (program 0113).
Contralateral electrode placement.
Intensity above threshold sensitive.
Application time: 1 - 2 minutes, max. 6 minutes, daily. Total number of treatments: 3 - 5.
54. Relaxatio mm. superficiales
Emitter head: 1 cm
2
. Frequency = 3 MHz, 1:1, 1:2, 1:4 PPR. Intensity: 0.5 - 1.0 W/cm
2
, step 0.1 W/cm
2
.
Electrotherapy: TENS cont., f = 100Hz (program 1725)
Contralateral electrode placement.
Intensity above threshold sensitive.
Application time: 1 - 10 minutes, step 1 minute, daily. Total number of treatments: 1 - 6.
55. Relaxatio mm. profundi
Emitter head: 4 cm
2
. Frequency = 1 MHz, continual PPR. Intensity: 0.5 - 1.0 W/cm
2
, step 0.1 W/cm
2
.
Electrotherapy: AMP, FM 100 - 200 Hz, steps 3/3 (program 0947)
Contralateral electrode placement.
Intensity at threshold motor.
Application time: 1 - 10 minutes, step 1 minute, daily. Total number of treatments: 1 - 6.
56. Trigger points superficialis
Emitter head: 1 cm
2
. Frequency = 3 MHz, 1:2 PPR. Intensity: 0.5 W/cm
2
.
Electrotherapy: TENS cont., f = 100 Hz const. (program 1625)
Contralateral electrode placement.
Intensity at threshold motor.
Application time: 1 minute, daily. Total number of treatments: 1 - 5.
57. Trigger points profundi
A/ Trigger point finding
Emitter head: 1 cm
2
(a 4 cm
2
head can also be used, but proper aim is difficult and degree of accuracy is
lowered). Frequency = 1 MHz, 1:4 PPR. Intensity: 2.0 W/cm
2
.
Electrotherapy: TENS continuous, imp. 0.05 ms, asymmetrically biphase, f = 100 Hz, (program 1525)
Procedure:
First apply contact gel (electrically conductive) to the examined area. Set the ultrasound parameters.
Outside the hyper-allergic zone (HAZ) or the assumed trigger (TP) set the intensity of the electrical to threshold
sensitive and through a fluent head movement we move to the assumed and reflective changed area. HAZ is
announced by the patient as a increased intensity of a feeling by electric current bypass. TP localisation can be
found in two ways:
1.
The patient reports a typical radiating pain.
2.
Asymptotic motor effect (muscle contraction) in the affected muscle.