BTL – 07p
operating manual & user's guide
page 10
atmospheric pressure:
650 hPa to 1100 hPa
Dimensions:
375 x 110 x 225 mm
Weight:
3.5 kg approx.
The unit is designed to comply with
standards
EN 60601-1:1990
EN 55 011
Equipment class:
class I
Type B applied part:
BF (symbol
)
Protection: IP
20
Mains transformer fuses used:
primary side (external):
2 x T 250 mA
secondary side (internal):
1 x F 100 mA, 2 x T 250 mA, 2 x T 630 mA
End of therapy indicator:
acoustic signal
ON indication:
POWER green control light
Proportional contact indication :
CONTACT green bar-graph
Error or bad skin contact indication:
ERROR and CONTACT red controls
ON/OFF switch marking:
0/1
Standard accessories:
1 ultrasound probe with area 1 cm
2
1 ultrasound probe with area 4 cm
2
1 bottle ultrasound gel
2 spare fuses (external)
Pulse width
-t
i
[ms] t
p
[ms]
T [ms]
f [Hz]
Duty factor
10 10 20 50 1:2
5 15 20
50
1:4
2.5 17.5 20 50 1:8
1.25 18.75 20 50 1:16
5 5 10
100
1:2
2.5 7.5 10 100 1:4
1.25 8.75 10 100 1:8
0.625 9.375 10 100 1:16
Important!
Each BTL-07p instrument has its own specific ultrasonic emitter heads that are produced and set only for that
particular unit. Do not use emitter heads from a different BTL-07p unit or an ultrasound unit made by another
producer. Protect the emitter heads from shock or intense cold. Do not bend the cable. Be especially careful
with the emitter head connection area. Regularly disinfect emitter heads with a commonly used disinfectant.