BTL 6000 SWT TOPLINE User Manual Download Page 31

 

MAINTENANCE AND SAFETY INSTRUCTIONS | PAGE 31 OF 38 

4.2  GENERAL SAFETY PRECAUTIONS    

 

Before turning on the device for the first time, read this manual carefully. 

 

The device may only be used under the supervision of the physician prescribed the treatment. 

 

All  personnel  must  be  properly  trained  before  using  the  device.  This  training  should  include  the  servicing  technique, 

maintenance, verification of proper functioning and safety precautions associated with using the device. 

 

Shockwave therapy requires direct bodily contact, therefore do not use it on damaged skin! 

 

Do not apply shockwave therapy:  To areas of the body with gas-containing organs, close to large nerve bundles 

or blood vessels, on the spinal cord or on the head! 

 

We do not recommend applying shockwave to the body parts under local anaesthesia.  

 

The device does not use any drugs, creams, gels or other substances which are an integral part or which are applied by 

its use.  

 

The device is equipped with a protection system that prevents the connection of accessories other than those supplied by 

the manufacturer. Consequently it will not operate with equipment from other manufacturers. 

 

Only  the  power  supply  adapter  approved  and  supplied  by  the  manufacturer,  complying  with  the  specifications  listed  in 

Chapter Technical Parameters, may be connected to the device. 

 

The product complies with Class A emissions according to CISPR11, IEC 60601-1-2. 

 

Portable and mobile high-frequency communication devices (such as mobile phones) may affect the function of the device. 

 

The electrical cables which are to be connected to the device must be installed and tested according to the existing valid 

standards  (IEC  364).  If  it  is  not  known  whether  cables  are  safe  and/or  correct,  then  they  should  be  checked  and/or 

upgrade by an inspection engineer.  

 

Check  whether  the  parameters  of  the  main  power  supply  correspond  to  the  requirements  of  the  device  according  to 

Technical Parameters section 

 

The device requires the environmental conditions that are stated in the Technical Parameters section. It must not be used 

in  an  environment  where  there  is  a  danger  of  explosion  or  of  water  penetrating  the  device.  The  device  cannot  be  in 

connection with flammable anaesthetics or oxidizing gasses (O2, N2O, etc.).  

 

Do not place the device in direct sunlight or near strong electromagnetic fields to prevent mutual functionality influence. If 

this  happens,  move  the  device  further  away  from  the  source  of  interference  or  contact  an  authorized  BTL  service 

department.  

 

Inspect the device thoroughly before each use. Look for loose cables, cracked cable insulation, cracks in the shockwave 

applicator’s housing and functional behavioural differences in the display or the operating elements.  If any anomalies or 

inconsistencies are found, stop using the device and contact an authorized BTL service department.  If the behaviour of 

the device shows any divergences from the functionality procedures described in this user’s manual, stop using the device 

and contact an authorized BTL service department. 

 

If the device shows any defects or if there are any doubts concerning its correct and safe functioning, terminate therapy 

immediately. If the source of the concern can be determine after a thorough study of the user's manual, then contact an 

authorized BTL service department immediately.  If the device is not used in accordance with this manual or if it is used 

when  the  device  exhibits  functional  differences  from  those  stated  in  this  manual,  then  BTL  is  not  responsible  for  any 

damage to or caused by the device. 

 

Do not try to open, remove protective covers, or dismantle the device for any reason. There is a danger of electrical shock 

and/or serious injury. Even the replacement of the lithium battery and the air micro-filter must be done an authorized BTL 

service department only!  

 

All  material  and  parts  which  come  into  direct  contact  with  the  client's  body  (such  as  cleaning  agents,  the  applicators, 

electrodes)  must  respectively  comply  with  the  standards  related  to  irritants,  allergens,  toxins,  geotaxis  and  carcinogens 

(ISO  10993-1,  ISO  10993-3,  ISO  10993  5).  The  user  is  responsible  for  all  these  materials  and  parts  if  they  were  not 

supplied by an authorized BTL equipment supplier. 

Summary of Contents for 6000 SWT TOPLINE

Page 1: ...GENERAL CHARACTERISTICS OF THE DEVICE PAGE 1 OF 38 BTL 6000 SWT TOPLINE USER S MANUAL v100AS08 01 2010EN...

Page 2: ...ince we believe an ongoing relationship with our customers is critically important to our future product line While we would like you to start using your new equipment right away we encourage a thorou...

Page 3: ...Therapy Parameters Via the diag Button 13 2 6 2 Setting Therapy Parameters Via the prog Button 13 2 6 3 Setting Therapy Parameters Manually User Setup Via the man Button 14 2 6 4 Setting the Number o...

Page 4: ...2 11 4 1 Applicator Button Mode 24 2 11 4 2 Applicator Kit Replacement Wizard 24 3 LIST OF STANDARD AND OPTIONAL ACCESSORIES 25 4 MAINTENANCE AND SAFETY INSTRUCTIONS 26 4 1 1 Shock Transmitter Replace...

Page 5: ...Topline is the predefined programs stored in the memory of the main unit Based on detailed research and practical use of the device the well organized predefined programs will provide recommendations...

Page 6: ...positive pressure pulse Its diffusion is only limited by the actual tissue absorption and eventual reflections on acoustic non homogeneities A shockwave is defined as a pressure pulse with these char...

Page 7: ...e tension Analgesic effect o Destruction of afferent nerves and nerve receptors o CNS stimulants sensed as pain are also inflammation transmitter substances o Regression of pain caused by local ischem...

Page 8: ...ade 1 Arthrosis Acupuncture Pain trigger points or painful points in muscles 1 8 CONTRA INDICATIONS FOR SHOCKWAVE TREATMENT Application to certain tissues The eyes and the surrounding area the myocard...

Page 9: ...t and stop therapy 6 ON OFF switch back lit in blue when the control unit is on 7 USB port in the space of the device s grip for use only in compliance with IEC 60950 1 The USB port serves only for se...

Page 10: ...00 SWT TOPLINE 14 connector for shockwave applicator 15 control unit mains fuse 16 connector for power cable 17 power on off switch 18 vessel for collecting condensed water 19 type label contains type...

Page 11: ...estions please contact an authorized service of BTL devices Procedure 1 First connect the device in mains by means of the supplied power supply adapter which you will connect to the connector on the r...

Page 12: ...ative purposes but others can be pressed and activated These basic elements include 3D buttons which can be pressed to change the indicated values Informative text Channels tabs to switch between the...

Page 13: ...AMETERS VIA THE DIAG BUTTON A list of diagnosis will display after pressing the diag button Each tab channel has its own assigned list of diagnosis To find a diagnosis quickly press the button for the...

Page 14: ...dual buttons will open an individual menu and pop up boxes for the settings Most of the pop ups are accompanied with illustrative pictures and or symbols 2 6 4 SETTING THE NUMBER OF SHOCKS The total n...

Page 15: ...cks 2 6 5 SETTING THERAPY INTENSITY The intensity power of the shockwave therapy can be set on the therapy parameters screen even during the course of therapy After pressing the intensity button it is...

Page 16: ...ameters are correctly set The device will warn of any possible discrepancies To interrupt therapy press the pause button on the screen or the start stop key on the panel of the device To resume the in...

Page 17: ...the treated tissue during therapy Minimizes the transmission of reverse shocks to the hand of the therapist due to the built in air damper An integral part of the applicator is the ergonomic hand rest...

Page 18: ...r These examples illustrate the application of shockwaves in different areas and several of ways of holding the applicator Painful shoulder Trigger points Tibial Edge Syndrome Calcar Calcanei Plantar...

Page 19: ...ed in both lists below the line After entering this data it is possible to assign a unique name to that particular therapy program The name of the therapy program will consist of the letter W for shoc...

Page 20: ...e of the recently used therapies on the selected tab After pressing the load button the therapy can be restarted and its parameters viewed 2 11 UNIT SETTINGS THE MENU BUTTON After pressing menu it is...

Page 21: ...d shocks etc 2 11 2 ENCYCLOPAEDIA The encyclopaedia provides information about diagnoses possible therapies and examples of how to use the device A hard copy of the encyclopaedia is always supplied wi...

Page 22: ...f standard sounds from the manufacturer for example the signaling the therapy operations All audio tones can be muted no sound or modified as required Allows the editing of individual audio schemes th...

Page 23: ...rences between the modes can be best seen on the therapy settings screen which will always appears before starting of therapy 2 11 3 9 Touch Panel Calibration If the buttons on the touch screen do not...

Page 24: ...e selected way of control it If the button is pressed for more than 1 second then after its subsequent release therapy will stop This will be the same in the continual mode See below During the time i...

Page 25: ...rom BTL For more detailed information on individual accessories see the enclosed leaflet Standard accessories 1x shockwave applicator with multi focus shock transmitter 15mm 1x replaceable multi focus...

Page 26: ...surface of the shock transmitter can be washed with warm water with its cap on However to completely remove all of the contact gel from the shock transmitter it will be necessary to unscrew the shock...

Page 27: ...ti focus shock transmitter 15mm 1x replaceable multi focus shock transmitter 9mm 1x replaceable focusing shock transmitter 15mm Replacement procedure 1 Unscrew the screw cap of the shock transmitter b...

Page 28: ...parts kit included in the BTL 6000 SWT Topline package Do not use damaged applicators There is a risk of injury to the operating staff or the client The spare parts kit includes 1x exchangeable kit 1x...

Page 29: ...STRUCTIONS PAGE 29 OF 38 3 Use the key to unscrew the screw cap of the cartridge See the pictures below 4 Remove the worn out part from the cartridge from the applicator 5 Remove the plastic cover fro...

Page 30: ...little bit inside the applicator case The replaced part must move freely 7 After inserting the replaced part firmly screw the screw cap of the cartridge back in place using the key 8 When replacing th...

Page 31: ...the Technical Parameters section It must not be used in an environment where there is a danger of explosion or of water penetrating the device The device cannot be in connection with flammable anaesth...

Page 32: ...paired or replaced by the user Do not remove the cover from the control unit All repairs must be done by an authorized BTL service department Do not disconnect an applicator during therapy 4 3 USED SY...

Page 33: ...127 2 Power switch according to IEC 60601 1 On the back of device positions 0 and I To disconnect from the mains unplug the male plug of the power supply adapter from the mains socket outlet Design We...

Page 34: ...529 IP20 Class according to IEC 60601 1 II Class according to MDD 93 42 EEC Ilb Type of operation permanent Operating conditions Ambient temperature 10 C to 40 C Relative humidity 30 to 75 Atmospheric...

Page 35: ...pline is suitable for use in the specified electromagnetic environment The purchaser or user BTL 6000 SWT Topline should assure that it is used in an electromagnetic environment as described below Imm...

Page 36: ...e 2 note These are guidelines Actual conditions may vary Recommended separation distance BTL 6000 SWT Topline is intended to be used in electromagnetic environment with controlled RF disturbances The...

Page 37: ...Testing and measurement techniques Electrical fast transient burst immunity test IEC 61000 4 4 am1 ed2 0 2009 EN ISO 61000 4 4 2005 Electromagnetic compatibility EMC Part 4 Testing and measurement tec...

Page 38: ...ransferred by any means including electronic mechanical and photographic without the prior written approval from BTL Industries Limited BTL Industries Limited operates on a policy of continuous develo...

Reviews: