Measurement range
Resolution
Accuracy
200 Ω
0.1 Ω
± 1 % + 5 digits
2 kΩ
1 Ω
± 1 % + 5 digits
20 kΩ
10 Ω
200 kΩ
100 Ω
2 MΩ
1 kΩ
20 MΩ
0.01 MΩ
± 1.8 % + 5 digits
-
Measuring voltage: 0.25 V.
Diode test
With the selector switch set to "
/ ◦))". Insert the black test lead into
the
‘COM’ socket and the red test lead into the V/Ω/TEMP/CAP socket.
Using the test probes, touch the test points of the test object. Red test
lead = anode, black test lead = cathode. The forward voltage drop is
displayed.
Measurement range
Resolution
Display
1 mV
Forward voltage
-
Forward current: approx. 25 µA, reverse voltage: approx. 2.8 V.
Continuity testing
With the selector switch set to "
/ ◦))". Insert the black test lead into
the
‘COM’ socket and the red test lead into the V/Ω/TEMP/CAP socket.
Using the test probes, touch the test points of the test circuit. An acous-
tic signal is emitted if a re
sistance under 70 Ω is measured.
Important: Isolate from the power supply and discharge ca-
pacitors in the circuit to be measured.
Measurement ran-
ge
Function
◦))
The integrated buzzer signals up to a resistance of
70
Ω