measuring range). Using the test probes, touch the test points of the
test object. Read measurement value on the display.
You must use the
‘10 A’ socket when measuring currents
above 200 mA.
Alternating current
Measurement
range
Resolution
Accuracy
2 mA
1 µA
± 1.2 % of r 5 digits
200 mA
0.1 mA
± 2.0 % of r 5 digits
10 A*
0.01 A
± 3.0 % of r 10 digits
Overload protection: mA range is protected with a 200 mA self-resetting
fuse.
-
10A range is protected with F 10 A / 1000 V.
-
In the 10A range observe the maximum operating times!
-
Frequency range: 40-400 Hz
*To protect against overheating, measurements should not
be taken for more than 10 seconds. Following this the de-
vice should be allowed to cool down for at least 15
minutes.