38
I-26/26R Incubator Shakers M1324-0050
User’s Guide
8
8
P
P
R
R
E
E
V
V
E
E
N
N
T
T
I
I
V
V
E
E
M
M
A
A
I
I
N
N
T
T
E
E
N
N
A
A
N
N
C
C
E
E
WARNING!
When performing maintenance on the unit, always turn off the shaker
and disconnect the power cord from the power supply.
8.1
Cleaning External Surfaces
The unit may be cleaned using a damp cloth or any standard household or laboratory
cleaner to wipe down its outer surfaces. Do not use abrasive or corrosive compounds to
clean this instrument, as they may damage the unit and void the warranty.
8.2 Fuse
Replacement
I-26 models require one 8-amp electrical fuse, which is housed in the fuse holder located
on the M1324-7004 PCB (
see Figure 15d
). I-26R models have two 8-amp electrical
fuses, housed in the two fuse holders located on the M1324-7003 PCB (
see Figure 15e
).
To check or replace a fuse:
1. Set
the
ON/OFF SWITCH
to Off and disconnect the
POWER CORD
from the power
source.
2. Remove the front right bezel from the shaker: with one hand, grasp the top edge, and
with the other, using the cut-out provided on the bottom, grasp the bottom edge. Pull
the bezel up from the bottom, then out and away. Set it aside.
3.
With reference to Figure 15a
, remove the two screws that fasten the electrical panel
in place. Set the screws aside for reuse.
Figure 15a: Front Bezel Removed
Remove
screws
ELECTRICAL
PANEL