BRUKER BIOSPIN
User Manual Version 002
General Hardware Setup
On SB probes the location may differ, but the connectors are (if present in the
case of PICS) easily identified by the type of connector.
Figure 3.14. Spin Rate Monitor Cable Connector for 2 Different Types of SB
Probes
Warning: The spin rate monitor cable has a probe side connector that is ex-
actly the same as the power supply cable for the B-TO thermocouple oven
used with many high resolution probes. If this cable is connected to the
spin rate monitor assembly at the probe, the latter will be destroyed. Make
sure the B-TO cable and the MAS cable (labelled “probe” at the probe side)
are labelled such that they cannot be mistaken!
MAS Tubing Connections
3.5
For any type of fast spinning probe, compressed gas is used to provide the spin-
ner bearing and drive gas. Please refer to the installation or site planning manuals
to learn about the gas requirements. The most important parameters are:
•
Mains pressure (should be at least maximum required pr1 bar, to pro-
vide pressure regulation range).
•
Bearing pressure: up to 4.5 bar (as of February 2008)
•
Drive pressure: up to 4.5 bar (as of February 2008)
This means that at least 5.5 bars of pressure should be available at the outlet. If
the pressure droop along the supply tube is substantial, the internal pressure may
drop below 5 bars, then the MAS unit stops to regulate and gives a warning.
Consequently, we recommend a primary (inlet) pressure of min. 6 bar, preferably
7-8 bar (max. 10 bar) and a low loss gas line (8mm inner diameter, distance
≤
5
meters) between the instrument and the gas supply, to assure trouble free opera-
tion even under conditions of high gas throughput. The maximum throughput de-
pends on the experimental conditions and the probe type.
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...