![Bruker BioSpin Solid State NMR User Manual Download Page 222](http://html1.mh-extra.com/html/bruker-biospin/solid-state-nmr/solid-state-nmr_user-manual_2821040222.webp)
222 (327)
BRUKER BIOSPIN
User Manual Version 002
Basic MQ-MAS
Data processing
17.4
Processing parameters should be set according to the following table:
Data obtained with
mp3qzfil.av
from nuclei with spin I = 3/2 can be processed with
xfb
, if
IN10
has been set appropriately to run a split-t
1
experiment. If this is not the
case data can be sheared in order to align the anisotropic axis along the F2 axis.
This is done with the AU program
xfshear
. The AU program checks the nucleus
to determine the spin quantum number, checks the name of the pulse program
and decides what type of experiment has been performed. In case the nucleus is
unknown to the program, or the pulse program has a name that does not contain
a string “nq” nor “nQ” (with n=3, 5, 7, 9), the required information is asked for by
the program, in order to calculate the shearing correctly.
Note that using a user designed pulse program that contains e.g. a string “5q” but
performs a 3Q experiment (and vice versa) will yield erroneous shearing. When
started the AU program prompts for “Apply ABS2?” and “F1 shift in ppm:”. It is ad-
visable to calculate a baseline correction after F2 Fourier transform. Note that the
range defined by ABSF1 and ABSF2 is used for this. You should make sure that
Table 17.4.
Processing Parameters for 2D FT
Parameter
Value
Comments
F2 (acquisition dimension)
SI
Usually set to one times zero filling.
WDW
no
Don’t use window function.
PH_mod
pk
Apply phase correction.
BC_mod
no
No DC correction is required after full phase cycle.
ABSF1
1000 ppm
Should be outside the observed spectral width.
ABSF2
-1000 ppm
Should be outside the observed spectral width.
STSR
0
Avoid strip FT.
STSI
0
Avoid strip FT.
TDoff
0
Avoid left shifts or right shifts before FT.
F1 (indirect dimension)
SI
256
Sufficient in most cases.
WDW
no
Don’t use window function, unless F1 FID is truncated.
PH_mod
pk
Apply phase correction.
BC_mod
no
No DC correction is required after full phase cycle.
ABSF1
1000
Should be outside the observed spectral width.
ABSF2
-1000
Should be outside the observed spectral width.
STSR
0
Avoid strip FT.
STSI
0
Avoid strip FT.
TDoff
0
Avoid left shifts or right shifts before FT.
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...