BRUKER BIOSPIN
User Manual Version 002
Basic MQ-MAS
Figure 17.4. Nutation profiles of selective and non-selective pulses.
Left diagram shows signal intensity of
87
Rb resonances in RbNO3 as a function of
a non-selective pulse at approx. 150 W RF power, the right diagram shows the
signal intensity as function of a selective pulse at less than approx. 0.5 W. Spectra
are taken on AV500WB at a Larmor frequency of 163.6 MHz with 2.5 mm CP/
MAS probe spinning at 25 kHz. Note the different scaling of x-axis, which is dis-
played as “ppm” but corresponds to the used pulse lengths in µs (apart from the
sign).
Once the central transition selective 90° pulse is calibrated the parameters can be
copied to a new data set with
iexpno
, and the MQMAS pulse program can be
loaded. Available pulse programs are
mp3qzqf
and
mp3qzfil
. The first is a 3-pulse
sequence, the second a 4-pulse sequence. The sequence with fewer pulses will
be slightly more sensitive, whilst the 4-pulse sequence can be used as an initial
set-up for experiments with sensitivity enhancement methods like DFS or FAM
(see
"MQ-MAS: Sensitivity Enhancement" on page 231
describing sensitivity
enhancement methods).
In
the starting parameters for the set-up are displayed. This table
gives typical values for the pulses and powers that should be close to the final val-
ues confirmed by the optimization procedure. Parameters like
O1
,
TD
,
SWH
,
RG
,
should already be set in the standard 1D spectrum. For 4 mm probes these pulse
lengths are about the limit of what can be achieved, for 2.5 mm probes somewhat
shorter pulses can be obtained. For I = 3/2 and I = 5/2 nuclei the ratio of p1/p2
≈
3.
For
pl11
an initial value that corresponds roughly to 300 W can be used. Optimi-
zation will be done on the first increment of the 2D sequence, i.e.
d0
= 1 µs. Two
strategies for the optimization procedure can be followed; either the pulse lengths
p1
and
p2
or the power level
pl11
can be optimized for maximum signal ampli-
tude. However, the latter can be disadvantageous because a power level above
the probe limit might be applied, in order to clearly determine the optimum power.
In the case of 300 W amplifiers the maximum signal amplitude may not be ob-
tained even at full power, with the chosen pulse lengths.
Summary of Contents for Solid State NMR
Page 8: ...8 BRUKER BIOSPIN User Manual Version 002 Contents...
Page 14: ...14 327 BRUKER BIOSPIN User Manual Version 002 Test Samples...
Page 118: ...118 327 BRUKER BIOSPIN User Manual Version 002 Basic CP MAS Experiments...
Page 122: ...122 327 BRUKER BIOSPIN User Manual Version 002 FSLG HETCOR Figure 8 3 The ased Display...
Page 154: ...154 327 BRUKER BIOSPIN User Manual Version 002 Proton Driven Spin Diffusion PDSD...
Page 178: ...178 327 BRUKER BIOSPIN User Manual Version 002 SUPER...
Page 192: ...192 327 BRUKER BIOSPIN User Manual Version 002 Symmetry Based Recoupling...
Page 212: ...212 327 BRUKER BIOSPIN User Manual Version 002 Relaxation Measurements...
Page 270: ...270 327 BRUKER BIOSPIN User Manual Version 002 Double CP...
Page 290: ...290 327 BRUKER BIOSPIN User Manual Version 002 Modified W PMLG...
Page 302: ...302 327 BRUKER BIOSPIN User Manual Version 002 CRAMPS 2D...
Page 304: ...304 327 BRUKER BIOSPIN User Manual Version 002...
Page 305: ......
Page 306: ......
Page 307: ......
Page 308: ......
Page 318: ...318 327 BRUKER BIOSPIN User Manual Version 002 Tables...
Page 326: ...326 327 BRUKER BIOSPIN User Manual Version 002 Index...
Page 327: ...User Manual Version 002 BRUKER BIOSPIN 327 327 End of Document...