As a result, freezing profiles are extremely consistent from one run to the next. Also, because of this low ther-
mal mass, the CoolCell SV10 container will not cause a rise in local freezer temperature and will protect nearby 
samples already stored in the freezer.  Low thermal mass also means CoolCell SV10 container will rapidly return 
to room temperature for another freezing cycle (see fast recycle instructions above).

CoolCell SV10 freezing performance

The CoolCell SV10 container will freeze six serum vials 
each containing 5ml of cell suspension at -1

o

C

 per minute 

when placed in a -8o

o

C

 environment (mechanical freezer 

or dry ice locker).   The graph (left) displays a typical 
freezing profile obtained using these conditions.  
Note:  “CC-12” refers to CoolCell 12-well 1ml cryovial 
freezer, #BCS-136.

Troubleshooting

Care and Cleaning

The CoolCell SV10 container  is constructed of closed cell crosslinked polyethylene foam and a solid thermo-
conductive core. CoolCell SV10 container is compatible with prolonged cryogenic temperature exposure.  The 
foam may be cleaned by water and mild soap.  Rinse and dry thoroughly. CoolCell SV10 container is resistant to 
alcohols and 10% bleach solutions. Do not autoclave. Maximum temperature exposure: 60

o

C

. Avoid prolonged 

exposure to UV light sources.

As a result, freezing profiles are extremely consistent from one run to the next. Also, because of this low ther-
mal mass, the CoolCell SV10 container will not cause a rise in local freezer temperature and will protect nearby 
samples already stored in the freezer.  Low thermal mass also means CoolCell SV10 container will rapidly return 
to room temperature for another freezing cycle (see fast recycle instructions above).

CoolCell SV10 freezing performance

The CoolCell SV10 container will freeze six serum vials 
each containing 5ml of cell suspension at -1

o

C

 per minute 

when placed in a -8o

o

C

 environment (mechanical freezer 

or dry ice locker).   The graph (left) displays a typical 
freezing profile obtained using these conditions.  
Note:  “CC-12” refers to CoolCell 12-well 1ml cryovial 
freezer, #BCS-136.

Troubleshooting

Care and Cleaning

The CoolCell SV10 container  is constructed of closed cell crosslinked polyethylene foam and a solid thermo-
conductive core. CoolCell SV10 container is compatible with prolonged cryogenic temperature exposure.  The 
foam may be cleaned by water and mild soap.  Rinse and dry thoroughly. CoolCell SV10 container is resistant to 
alcohols and 10% bleach solutions. Do not autoclave. Maximum temperature exposure: 60

o

C

. Avoid prolonged 

exposure to UV light sources.

Problem

Solution

Vials do not freely fit in the 

chambers

The CoolCell SV10 container is designed to fit standard serum vial 

shaped vessels up to 23mm in diameter and up to 47 mm in height.  

Check that flag style labels, if used, will not bind and hinder insertion 

or removal.

Vials are stuck in CoolCell SV10 

after freezing

It is likely moisture was in the vial chambers or on the sample vial prior 

to freezing.  Remove the core (black ring) and tap the CoolCell SV10 to 

dislodge vials.

The lid does not fully seat

Ensure that sample vials are properly seated.  The maximum height of 

the tube is 47mm.

Problem

Solution

Vials do not freely fit in the 

chambers

The CoolCell SV10 container is designed to fit standard serum vial 

shaped vessels up to 23mm in diameter and up to 47 mm in height.  

Check that flag style labels, if used, will not bind and hinder insertion 

or removal.

Vials are stuck in CoolCell SV10 

after freezing

It is likely moisture was in the vial chambers or on the sample vial prior 

to freezing.  Remove the core (black ring) and tap the CoolCell SV10 to 

dislodge vials.

The lid does not fully seat

Ensure that sample vials are properly seated.  The maximum height of 

the tube is 47mm.

15 Elizabeth Dr., Chelmsford, MA 01824 USA  brookslifesciences.com/contact-us  brookslifesciences.com

CoolCell SV10 Instructions, May 2018

CoolCell SV10 ©Brooks Automation, Inc. All rights reserved.

15 Elizabeth Dr., Chelmsford, MA 01824 USA  brookslifesciences.com/contact-us  brookslifesciences.com

CoolCell SV10 Instructions, May 2018

CoolCell SV10 ©Brooks Automation, Inc. All rights reserved.

Reviews: