1.5 Hook-up with 6DIN Connector, 6-pin round
See hook-up document no. 9.16.089 of the shipment’s accompanying documentation.
Mass flow meter D-62
0:
1 + power
4
+ output signal
2 0 power
5
n/c
3 0 output signal
6
n/c
Mass flow controller D-62
1 / D-62
3:
1 + power
4
+ output signal
2 0 power
5
+ setpoint signal
3 0 output signal
6
n/c
1.6 Hook-up with 15-pin sub-D Connector
Typ M for mass flow controller D-62
1 / D-62
3:
1
n/c
9
n/c
2
+ output signal
10 n/c
3
n/c
11 n/c
4
n/c
12 0 output signal
5
0 power
13 n/c
6
n/c
14 n/c
7
+ power
15 shield
8
+ setpoint signal
Typ B for mass flow controller D-62
1 / D-62
3:
1
0 power
9
0 power
2
+ output signal
10 0 power
3
0 power
11 n/c
4
n/c
12 n/c
5
+ power
13 n/c
6
n/c
14 shield
7
n/c
15 n/c
8
+ setpoint signal
Please pay attention to the label with the hook-up fitted at the cable.
Be sure that the integrity of the shielding is not affected when connecting the system to
other devices (e.g. to PLC). Only shielded wire terminals can be used.