Bronkhorst®
Instruction Manual ES-FLOW™
9.17.145E
23
The
Counter
sub menu provides the following parameters and functions:
Parameter
Description
Supported values
Counter mode
Selected counter mode
·
off
·
up to limit
·
up
Counter reset
Reset method
·
automatic
·
manual
Type*
Measurement type
Note:
changing this setting will reset both the counter and
the totalizer
·
volume
·
mass
Counter unit*
Measurement unit for the counter
Depending on selected
measurement type
Counter limit
Counter limit or batch size
0…9999999.999
Totalizer unit*
Measurement unit for the totalizer
Depending on selected
measurement type
Setpoint change
Specifies whether or not to change setpoint after reaching
counter limit
·
yes
·
no
Setpoint
New setpoint after reaching counter limit (until counter
reset)
Setpoint range
Overrun correction*
Prevents batch size overshoot by stopping flow before
reaching configured batch size (
Counter limit
)
·
higher value: faster correction, less dosing cycles needed
·
lower value: slower correction, more dosing cycles needed
0.000…1.000
Controller gain*
Prevents batch size overshoot by reducing flow when
approaching configured batch size (
Counter limit
)
·
default value: 10
·
higher value: shorter dosing time, more overshoot
·
lower value: longer dosing time, less overshoot
0 (disabled)
1.000…99999.999
Reset totalizer*
Resets the totalizer
*) Not available on ES-FLOW models before Mk II.