Brodie BiRotor Plus Installation & Operation Manual Download Page 12

Page 12/26

IOM Single Case BiRotor Plus R06

Installation

8.1 Installation

1.  The instrument should be mounted on a secure  

foundation. Provisions should be made for verti-
cally mounting the meter to ensure stability.

2.  The process piping should not place any undue 

strain on the instrument.

3.  Precautions should be taken to ensure that ther-

mal fl uid explansion does not raise the line pres-
sure above the maximum working pressure of the 
instrument.

4.  Process piping must be clean and free of any for-

eign matter.

5.  A strainer should be installed upstream of the  

instrument.

6.  An air eliminator should be installed upstream of 

the instrument if the process fl uid is expected to 
contain entrained air.

7. A 

fl ow limiting valve should be installed down-

stream of the instrument, this will maintain a back 
pressure on the instrument and prevent excessive 
fl ow rates.

8.  Isolation valves should be located at either ends 

of the instrument run and a bypass section in-
stalled. This will facilitate ease of component re-
moval, when required, and reduce loss of product.

8.2 Installing the Instrument

1.  Remove the inlet and outlet protection covers.

2.  Install the instrument into the pipe work using 

suitable hardware as specifi ed in the local codes 
and regulations. Ensure that the connections are 
made tight and torqued to the correct values.

3.  Connect the instrument wiring, refer to Figures 9-1 

and 9-2. Cable entry into the electrical enclosure 
is by two 3/4-14 NPT female threads.

4.  Use wiring appropriate for the location and  

operating conditions. Wiring glands and/or  
conduits must conform to local electrical codes 
and regulations if wiring is to be installed in a  
hazardous area.

NOTE:

For additional requirements on installation please 
refer to Chapter 2, Essential Instructions, at the 
beginning of this manual.

If the instrument is being installed with additonal 
accessories, the instructions for those accessories 
should be read and understood before continuing with 
the installation. The output signal from the pream-
plifi er, if one is fi tted, can be altered to interface with 
most electronic accessories. This is accomplished by 
the use of jumpers on the circuit board. Refer to  
Figure 8-1 for their confi gurations.

The UMB housing may be rotated in 90° increments 
to accomodate the mounting of accessories. If the 
UMB housing is removed for any reason, care must 
be taken to disconnect the sensor wires to prevent 
sensor damage. It is essential that the four retaining 
screws used to secure the UMB housing to the meter 
body be torqued to 55 in/lbs. 

     

Figure 8.1 Wiring connections

Summary of Contents for BiRotor Plus

Page 1: ...Page 1 26 IOM Single Case BiRotor Plus R06 Installation Operation Manual BiRotor Plus 2 Single Case...

Page 2: ...Page 2 26 IOM Single Case BiRotor Plus R06...

Page 3: ...General 10 2 Removal Replacement of Circuit Board 10 3 Removal Replacement of Pick Off Sensors 10 4 Removal of the Measuring Unit from the Process Line 10 5 Removal of the Measuring Unit from the Inst...

Page 4: ...ontents of this publication are presented for informational purposes only and while every effort has been made to ensure their accuracy they are not to be construed as warranties or guarantees express...

Page 5: ...ed with in the instruction manual or marked on the instruments name plates Install the equipment as specified in the installation instructions of the appropriate manual and in accordance to local and...

Page 6: ...custody transfer label this label details the working flow range associated with a particular weights and measures approval It should be noted that these may not be the same therefore in trade applic...

Page 7: ...e or uncontrolled chemical reactions It is the end users responsibility to provide fire protection measures and equipment in accordance with the local regulations It is the end users responsibility to...

Page 8: ...ust be made from stainless steel grade A1 70 or A2 70 and be torqued to a value of 55 in lbs upon installation its is the end users responsibility to ensure this happens It is the end users responsibi...

Page 9: ...cing the instrument upright on the inlet flange will aid drainage Process connections must be sealed to prevent leakage of residual product during shipment Contact the local carrier for information on...

Page 10: ...ting parts Principle of Operation The operation of the meter is embodied in the function of the measuring rotors they are always dynamically balanced but hydraulically unbalanced during operation The...

Page 11: ...y Voltage Less 5 70 mA max Open Collector Max voltage 30 VDC Max current 125 mA Max power 0 5 W Performance SB25X Linearity Standard Rotors 0 15 Over Standard Flow Range 0 25 Over Extended Flow Range...

Page 12: ...suitable hardware as specified in the local codes and regulations Ensure that the connections are made tight and torqued to the correct values 3 Connect the instrument wiring refer to Figures 9 1 and...

Page 13: ...must be closed all electrical connections are complete and all covers are in place 1 To pressurize the system slowly open the inlet valve so as to prevent system shock Slowly allow product to enter t...

Page 14: ...low the temperature to stabilize 9 3 Custody Transfer A meter factor will need to be established under actu al operating conditions if the instrument is to be used in custody transfer applications Thi...

Page 15: ...l measuring element contains closely meshed moving parts care should be taken not to insert figures into the rotors or timing gears as this will cause injury NOTE All Item Numbers referenced in this s...

Page 16: ...sor in the sensor housing If only one pick off sensor is present it should be inserted in the hole labelled A Secure with the hold down washer and Allen screws and replace the circuit board 10 4 Remov...

Page 17: ...rew from the end plate at the timing gear end of the measuring unit assembly NOTE A flat head screwdriver may be used in conjunction with the slots on the end plate to aid in its removal Excessive for...

Page 18: ...ing unit housing on it s side so that the name plate label in on the bottom and the outlet end facing away from you Then with the tapered ands of the rotors also facing away from you the right hand ro...

Page 19: ...ht hand timing gear to a torque setting of 35 in lbs 3 Loosen the nut of the left hand timing gear 4 Place 0 003 shims in front and behind the tooth of the left hand rotor Once the shims are in place...

Page 20: ...information consult factory Whenordering the following information must be furnished 1 Part number and description 2 Model number of flow meter 3 Serial number of flow meter 4 Quantity required When o...

Page 21: ...Page 21 26 IOM Single Case BiRotor Plus R06 Figure 11 1 Complete Measuring Unit Assembly...

Page 22: ...150 ANSI MTR s 1 56819 100N Outlet Housing 150 ANSI MTR s and NACE 56819 040M Outlet Housing DIN PN 16 and PN 40 MTR s 56819040N Outlet Housing DIN PN 16 and PN 40 MTR s and NACE 56819 300M Outlet Hou...

Page 23: ...1500909D Approved Stopping Plug 1 43 SD442 Plastic Cap Plug 1 Table 11 1 Parts List Single Case BiRotor Plus Continues Notes Items marked with a are recommended spare parts 1 These items are only ava...

Page 24: ...ction 7 Specifications Power failure Ensure power is connected to the device and all associated accessories Meter rotors jammed with debris Remove debris from rotors and check for damage to rotors tim...

Page 25: ...ie are not covered by this limited warranty and shall be at Buyer s expense Brodie shall not be obligated to pay any costs or charges incurred by Buyer or any other party except as may be agreed upon...

Page 26: ...m or cause of action Buyer agrees that in no event shall Brodie s liability to buyer and or its customers extend to include incidental consequential or punitive damages The term consequential damages...

Reviews: