background image

Document # 2103017 RevA 

Printed in USA © 2017

Table raises slightly slower than normal 

while the up pedal is held. Does one 

actuator continue to hum?

No

                                                         

Yes

Replace back actuator.

Return to step 1

Return to step 1

Call Brewer Tech Support 888-Brewer-1.

Test complete

Step 2.2 continue here

Step #

2.2

a

b

c

d

e

f

Table is low and does not rise. 

Do motors hum when up pedal is 

pressed?

No                                                   Yes

Table does not go all the way down. 

After table stops in the down direction, 

does continuing to press down on 

the foot control pedal cause motor 

operation, but no table movement?

No                                                   Yes

One lift actuator has failed. Replace 

both lift actuators. 

Return to step 1.

No backrest movement (powered back 

models). Does motor hum in the up 

direction, but there is no movement?

No                                                   Yes

Backrest or lift stuck in the full up 

position. No motor hum. Do control 

box relays click in both up and down 

direction?

No                                                   Yes

Backrest safety switch, 1 or more of 

4 is not returning to the closed posi-

tion. Support seat. Remove shroud to 

access control box and wiring area. 

Separate the Molex connector with 

orange, white, violet and black wires 

(power back) or orange and white 

wires on one side (manual back). On 

the connector half that has the thumb 

catch check continuity between the 

violet and black (backrest) and then 

the orange and white (seat lift). If no 

continuity on either circuit, replace 

with backrest safety switch kit.

Problem remains                       Fixed

Are there other problems not yet 

identified?

Yes                                                   No

6000 SERIES TABLE TROUBLESHOOTING CHART (continued)

Summary of Contents for access high-low

Page 1: ...ument 2103017 RevA Printed in USA 2017 Troubleshooting Guide The Brewer Company LLC N88 W13901 Main Street Suite 100 Menomonee Falls WI 53051 P 1 888 Brewer 1 F 262 251 2332 www brewercompany com Brew...

Page 2: ...ntacting the back cabinet 2 Press the BACK DOWN pedal The back should go to the full recline position which is level 3 Press the BACK UP pedal The back should go to the full up position 4 Press the BA...

Page 3: ...f 6 not closing when pressed Press each one and listen for click when pressed and again when released Replace as necessary Fixed Problem remains Seat back switches 1 or more of 4 not opening when pres...

Page 4: ...r 5 minutes with switch in either position Is warmer warming up Yes No Check for loose wire connections at terminal block s if S N is less than HL002100 or if S N has a C at the end Check that all plu...

Page 5: ...ckrest movement powered back models Does motor hum in the up direction but there is no movement No Yes Backrest or lift stuck in the full up position No motor hum Do control box relays click in both u...

Page 6: ...luminate Yes No Replace warmer outlet fuse s at back of table on plate where foot control and power cord plug in Does outlet operate Yes No Replace warmer switch Does warmer switch illuminate Yes No R...

Page 7: ...ment not stopped by stop button or foot control pedals No Yes The backrest actuator return chair that protrudes from the center of the back of the seat section is defective or its wiring harness blue...