Document # 2103017 RevA
3
Printed in USA © 2017
Complete the “
Operating Test
”.
Test Failed Test Passed
Test complete
Call Brewer Tech Support 888-Brewer-1.
Call Brewer Tech Support 888-Brewer-1.
No movement
Go to step 2.1
Some movement
Go to step 2.2
Start
Step #
* No relay clicking in control box. Table then functions after about 30 minutes.
1
2
3
4
5
6
7
Do all powered functions work?
Yes No
Go to step 5.1
for drawer warmer
Go to step 6.1
for outlet.
Return to step 1
Go to step 7.1
Do the 6 front limit switches work
properly?
Yes No
Front cabinet switches; 1 or more of 6
not closing when pressed. Press each
one and listen for click when pressed
and again when released. Replace as
necessary.
Fixed Problem remains
Seat back switches; 1 or more of 4 not
opening when pressed. Press each
one and listen for click when pressed
and again when released. Replace as
necessary.
Fixed Problem remains
Do the 4 seat back limit switches work
properly?
Yes No
Does the drawer warmer heat prop-
erly, and does the outlet work?
Yes or N/A No
Does the table stop operating after a
period of heavy use? *see below
No Yes
Control box is over temperature. Allow
30 minute cool down. Control box will
reset automatically.
Is there excessive noise, or are there
other problems which have not yet
been identified?
No Yes
6000 SERIES TABLE TROUBLESHOOTING CHART