Troubleshooting Guide

The Brewer Company, LLC

N88 W13901 Main Street, Suite 100

Menomonee Falls, WI 53051

P  1.888.Brewer.1

F  262.251.2332

www.brewercompany.com

Brewer

Summary of Contents for access high-low

Page 1: ...ument 2103017 RevA Printed in USA 2017 Troubleshooting Guide The Brewer Company LLC N88 W13901 Main Street Suite 100 Menomonee Falls WI 53051 P 1 888 Brewer 1 F 262 251 2332 www brewercompany com Brew...

Page 2: ...ntacting the back cabinet 2 Press the BACK DOWN pedal The back should go to the full recline position which is level 3 Press the BACK UP pedal The back should go to the full up position 4 Press the BA...

Page 3: ...f 6 not closing when pressed Press each one and listen for click when pressed and again when released Replace as necessary Fixed Problem remains Seat back switches 1 or more of 4 not opening when pres...

Page 4: ...r 5 minutes with switch in either position Is warmer warming up Yes No Check for loose wire connections at terminal block s if S N is less than HL002100 or if S N has a C at the end Check that all plu...

Page 5: ...ckrest movement powered back models Does motor hum in the up direction but there is no movement No Yes Backrest or lift stuck in the full up position No motor hum Do control box relays click in both u...

Page 6: ...luminate Yes No Replace warmer outlet fuse s at back of table on plate where foot control and power cord plug in Does outlet operate Yes No Replace warmer switch Does warmer switch illuminate Yes No R...

Page 7: ...ment not stopped by stop button or foot control pedals No Yes The backrest actuator return chair that protrudes from the center of the back of the seat section is defective or its wiring harness blue...