54
SB
I-240
Service Manual
11.
When the stable and reasonable data corresponding to the standard weight
has been received, the indicator will automatically go to Step8. Otherwise, it
will display
CAL.Er
and return to the previous steps.
5.4 Geographical Adjustment
1.
When
GEO
is selected, press the
TARE/PRESET
key to enter Geographical
Adjustment.
2.
When
CODE
is displayed, press the
key to select geographical
position code or input user local gravity value directly.
3.
When
CODE
is selected, select the position code of the scale being used (00-
70) according to the elevation and latitude from
by using the
key.
4.
Press
TARE/PRESET
key to confirm.
CAUTION! Only an authorized manufacturer representative or certified
verification personnel may make these changes. Changing the Geographical
setting alters the calibration values!