can press 

ON/OFF/ZERO

 to exit the calibration mode. Or load 5%-100%FS 

weight, and use 

PRINT/HOLD

 

and 

UNIT

 keys to input the loaded weight. 

If the input data is larger than 199999, it will show as “Hxxx” and “Lxxx”. If 
the triangular symbol on the left bottom of LCD window appears, it means 
that the digit being changed is the displayed most significant bit which can 
only be 0/blank or 1. Press 

TARE

 key to confirm your setting and the 

indicator will flash the input standard weight. Wait till the scale comes steady, 
and input A/D data as per the standard weight.   
The indicator will automatically go to next step, if the second point can be 
calibrated correctly. If there’s an error occurred, the scale will display “CAL. 
Er” and return back to step 5 for re-calibration. 

7.

 

When ‘CAL.P2’ is displayed, the scale will be calibrated on third calibration 
point. When xxxxxx kg (or lb) is displayed (100% FS is default), you can 
press 

ON/OFF/ZERO

 to exit the calibration mode or Place a standard 

weight (must be in the range of 10%-100% FS, and equal or larger than that 
for the second calibration point; this is also the range of your input number) 
on the scale. Use 

PRINT/HOLD

 and 

UNIT

 key to input the standard 

weight’s value. Use 

TARE

 key to confirm the standard weight and input 

number are correct. If the calibration weight for third point is same with that 
for second point and the calibration weight is more than 10%FS, input the 
standard calibration weight same as second point calibration and press 

TARE

 

key to confirm the setting. The indicator will flash the input weight. If the 
indicator get reasonable data (the input weight is correct, and the calibration 
weight of third calibration is more than equal to the calibration weight of 
second calibration), it will go to next step automatically. If there’s an error 
occurred, the scale will display “CAL.Er” and return back to step 5 for 
re-calibration. 

8.

 

When ‘CAL.P0’ is shown again, that means the scale will calibrate scale’s 
zero-point again. Now, you can press 

ON/OFF/ZERO

 to exit the calibration 

mode; or Move away any weight on the scale, press 

TARE

 key to confirm; 

the displayed data will blink. If the indicator gets reasonable data, it will 
calculate and store all parameters in EEPROM. And then it will auto-reset 
and display all segments of LCD, full capacity… like power on again. If 
there’s an error occurred in calibration, the scale will display “CAL.Er” and 

 

try to repeat from step 5. The scale will return to normal weighing mode.   

Attention: To ensure the accuracy of the scale, it is recommended to use over 

75% full capacity weight to calibrate. 

 

. DISPLAY ADC CODE OR INPUT WORKING VOLTAGE VALUE   

1.

 

In normal weighing mode, press and hold down 

ON/OFF/ZERO

 and 

HOLD/PRINT

 key more than 3s, until ‘

codE

’ is shown, this means you have 

been in display inner code mode; but first, the indicator will show the firmware 
version “xx.xx.xx”. In this mode, you can examine the inner working voltage, 
the stability of weighing system, the variety value of A/D data as per the loaded 
weight.

 

 

NOTE:  
1)

 

The increment of A/D code for FS weight must be larger or equal to 2 times 
of selected display division-n; otherwise, the calibration cannot be properly 
completed. Eg. The display division is 0.1kg. Load 100kg standard weight 
on the platform, the increment of A/D code is at least 2x 100kg/0.1kg= 
2x1000=2000. In this case, the scale can be calibrated. Otherwise, smaller 
division needs to be chosen.   

2)

 

The data should be stable; otherwise, the calibration cannot properly 
complete.  

2.

 

In this mode, you can calculate the proper ADC data at zero point by examining 
the A/D data for loaded weight. If the ADC increase for full capacity is NFS, 
the power-on zero range is set to Zp% FS (P12 setting) and zero key range is 
set to Zk% FS (P13 setting). Then proper ADC data of zero point is larger than 
(Zp%+ Zk%) x NFS.   
ADC increase for full capacity (NFS) can be making out by: Load the weight 
W on the platform, and the ADC increase for W weight is Nw. The ADC 
increase for full capacity WFS is (NFS)= (Nw)x (WFS)/W .   
Negative value may be displayed because of error connect of loadcell or error 
position of the zero-point potentiometer on PCB; however, the software only 
deals with positive value. So, if  the position of zero-point potentiometer is 
error, adjust potentiometer’s position to make the ADC data will be positive 
value and larger than (Zp%+ Zk%) x NFS. Normally the indicator is 
factory-calibrated, and end users do not need this operation. 

Summary of Contents for MS140/300

Page 1: ...MS140 300 Operation Manual Contents subject to change without notice Version 2 1 2012 01 10...

Page 2: ...aux appareils num riques de la Classe A prescrites dans le R glement sur le brouillage radio lectrique edict par le minist re des Communications du Canada Warnings Electrical installation For your pro...

Page 3: ...c device These devices emit RF and can cause unstable scale readings If your scale ever performs poorly try moving the scale to a different room or location Avoid using in condition of heavy vibration...

Page 4: ...tivity 0 3mV V 3mV V must be fit to 0 2uV display grad 2 Input Resistor 80 3 Output Resistor 10 K LOADCELL WIRING PIN 1 RED EXCITATION PIN2 BLACK EXCITATION PIN3 GREEN SIGNAL PIN4 WHITE SIGNAL 3 KEY F...

Page 5: ...ormal weighing mode and hold down time longer than 3s 1 Same meaning with that in 3 1 2 Same meaning with that in 3 1 3 Same meaning with that in 3 1 4 Power off the scale UNIT TARE ZERO HOLD PRINT UN...

Page 6: ...oint dot in calibration unit P9 and capacity s unit in calibration P10 If the setting of FS is more than 199999 regardless of decimal point and weight unit the FS capacity will be shown by first four...

Page 7: ...d data will blink If the indicator gets reasonable data it will calculate and store all parameters in EEPROM And then it will auto reset and display all segments of LCD full capacity like power on aga...

Page 8: ...Detailed setup 5 1 P1 xy Auto off time xy 00 15 xy 00 means no auto off function xy 01 15 means the scale turns off to save power after 01 15 minutes when no variety of weight on scale and no key ope...

Page 9: ...put compatible 8 data to the NCI SP1 format 5 5 P5 x Baud rate of RS232 communication x 0 1200bps x 1 2400bps x 2 4800bps x 3 9600bps x 4 19200bps 5 6 P6 x RS232 communication data format x 0 8N1 8 di...

Page 10: ...lb Not available 0 0002 lb 0 0001 kg 0 0002 lb Not available 0 002 lb 0 001 kg 0 002 lb Not available 0 02 lb 0 01 kg 0 02 lb 0 5 oz 15 0 2 lb 0 1 kg 0 2 lb 5 oz 2 lb 1 kg 2 lb Not available 20 lb 10...

Page 11: ...power off zero point as power on zero point and power off tare weight as current tare weight 5 16 P16 x Zero tracking range x 0 0d means no tracking x 1 0 25d x 2 0 5d x 3 1d x 4 1 5d x 5 2d x 6 3d x...

Page 12: ...f the zero point is error it will return nine _ characters b The character will be for negative weight or a space character for positive 19 weight Minus sign follows after the first digit c Useless le...

Page 13: ...ip of capacity and the setting of P7 P8 and P9 Table6 Capacity unit is kg or lb Resolution set by P7 Division set by P8 and P9 P8 0 0 0001 0 001 0 01 0 1 1 10 500 0 0500 0 500 5 00 50 0 500 5000 600 0...

Page 14: ...160000 1600000 100000 20 0000 200 000 2000 00 20000 0 200000 2000000 23 Table8 Capacity unit is kg or lb Resolution set by P7 Division set by P8 and P9 P8 2 0 0005 0 005 0 05 0 5 5 50 500 0 2500 2 50...

Page 15: ...as following drawings 1 Placed vertically 2 Placed horizontally 25 Key Definition summary KEY MODE DEFINITION HOLD PRINT Normal weighing mode Enter or exit HOLD mode output the data as per P4 P5 P6 s...

Page 16: ...x 0 1 2 3 4 3 Baud rate for RS232 0 1200bps 1 2400bps 2 4800bps 3 9600bps 4 19200bps 27 P6 x 0 1 2 0 RS232 format 0 8N1 1 7O1 2 7E1 P7 xy 00 31 13 Resolution select 500 600 750 800 1000 1200 1500 200...

Page 17: ...play 0 P16 x 0 1 2 3 4 5 6 7 8 8 Zero tracking range 0 0d no tracking 1 0 25d 2 0 5d 3 1d 4 1 5d 5 2d 6 3d 7 4d 8 5d P17 x 0 1 2 3 2 Data filter intensity 0 very weak 1 weak 2 middle 3 strong P18 x 0...

Page 18: ...845 246 6717 Fax 44 0 845 246 6718 Email sales brecknellscales co uk Web site www brecknellscales co uk USA Brecknell 1000 Armstrong Drive Fairmont MN 56031 Toll Free 800 637 0529 Tel 507 238 8702 Fa...

Reviews: